Literature DB >> 14739310

Björk-Shiley convexoconcave valves: susceptibility artifacts at brain MR imaging and mechanical valve fractures.

Maarten J van Gorp1, Yolanda van der Graaf, Bas A J M de Mol, Chris J G Bakker, Theo D Witkamp, Lino M P Ramos, Willem P T M Mali.   

Abstract

PURPOSE: To assess the relationship between heart valve history and susceptibility artifacts at magnetic resonance (MR) imaging of the brain in patients with Björk-Shiley convexoconcave (BSCC) valves.
MATERIALS AND METHODS: MR images of the brain were obtained in 58 patients with prosthetic heart valves: 20 patients had BSCC valve replacements, and 38 had other types of heart valves. Two experienced neuroradiologists determined the presence or absence of susceptibility artifacts in a consensus reading. Artifacts were defined as characteristic black spots that were visible on T2*-weighted gradient-echo MR images. The statuses of the 20 explanted BSCC valves-specifically, whether they were intact or had an outlet strut fracture (OSF) or a single-leg fracture (SLF)-had been determined earlier. Number of artifacts seen at brain MR imaging was correlated with explanted valve status, and differences were analyzed with nonparametric statistical tests.
RESULTS: Significantly more patients with BSCC valves (17 [85%] of 20 patients) than patients with other types of prosthetic valves (18 [47%] of 38 patients) had susceptibility artifacts at MR imaging (P =.005). BSCC valve OSFs were associated with a significantly higher number of artifacts than were intact BSCC valves (P =.01). No significant relationship between SLF and number of artifacts was observed.
CONCLUSION: Susceptibility artifacts at brain MR imaging are not restricted to patients with BSCC valves. These artifacts can be seen on images obtained in patients with various other types of fractured and intact prosthetic heart valves. Copyright RSNA, 2004

Entities:  

Mesh:

Year:  2004        PMID: 14739310     DOI: 10.1148/radiol.2303021619

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  Imaging and clinical characteristics of children with multiple foci of microsusceptibility changes in the brain on susceptibility-weighted MRI.

Authors:  Tetsu Niwa; Noriko Aida; Taro Takahara; Thomas C Kwee; Kazutoshi Fujita; Ayako Shishikura; Daiki Miyata; Tomio Inoue
Journal:  Pediatr Radiol       Date:  2010-04-28

2.  Prevalence and Evolution of Susceptibility-Weighted Imaging Lesions in Patients With Artificial Heart Valves.

Authors:  Philipe S Breiding; Jana T Duerrenmatt; Felix G Meinel; Thierry Carrel; Florian Schönhoff; Felix Zibold; Johannes Kaesmacher; Jan Gralla; Thomas Pilgrim; Simon Jung; Urs Fischer; Marcel Arnold; Thomas R Meinel
Journal:  J Am Heart Assoc       Date:  2019-08-05       Impact factor: 5.501

  2 in total