Measurement of hydration in the stratum corneum with the MoistureMeter and comparison with the Corneometer.

Esko Alanen1, Jouni Nuutinen, Kirsi Nicklén, Tapani Lahtinen, Jukka Mönkkönen.   

Abstract

BACKGROUND/
PURPOSE: MoistureMeter is a novel capacitive device for measuring the hydration of stratum corneum (SC). The capacitor 'plates' are formed by the probe material and the well-conducting epidermal-dermal skin layers, while the dry layer of SC acts as an 'insulator' of the capacitor. Due to this measurement principle, the measurement depth is not constant, but equal to the thickness of the dry layer of SC. The present study was undertaken to test the MoistureMeter SC-2 in an experimental setup simulating the uppermost skin structure. The sensitivity of the MoistureMeter was compared with a conventional device Corneometer CM 820 in eight healthy volunteers.
METHODS: The experimental setup consisted of a polyethene foil simulating SC and a mixture of cellulose and saline simulating the underlying skin layers with a high water content. The hydration of SC of volar forearm was measured in eight healthy volunteers both with the MoistureMeter SC-2 and Corneometer CM 820 after the application of three different skin formulations: base, base plus salt (2 wt% NaCl), base plus salt and glycerin (5 wt%). In the sorption-desorption test, the skin was wetted with a drop of water and the dehydration was followed with both devices for 2 min.
RESULTS: The test with the experimental setup agreed with the results predicted by the mathematical capacitance model. The hydration values of the MoistureMeter SC-2 and Corneometer CM 820 correlated well (R=0.75), but the relative range of readings with the MoistureMeter was approximately three times larger than with Corneometer. The MoistureMeter was insensitive for the salt (2 wt% NaCl) of the formulation and differentiated the hydration effect of glycerin better than the Corneometer. In the sorption-desorption test, the MoistureMeter demonstrated the individual differences between the volunteers, whereas with the Corneometer the differences were minimal.
CONCLUSION: The capacitive measuring principle of the MoistureMeter was demonstrated in an experimental arrangement. Although both instruments yielded equivalent data with the base formulation, the MoistureMeter was a more sensitive device than the Corneometer CM 820 and independent of added salt in the formulations.

Entities:  

Mesh:

Year:  2004        PMID: 14731246     DOI: 10.1111/j.1600-0846.2004.00050.x

Source DB:  PubMed          Journal:  Skin Res Technol        ISSN: 0909-752X            Impact factor:   2.365


  17 in total

1.  Subepidermal moisture surrounding pressure ulcers in persons with a spinal cord injury: a pilot study.

Authors:  Jeffrey John Harrow; Harvey N Mayrovitz
Journal:  J Spinal Cord Med       Date:  2014-11       Impact factor: 1.985

2.  Subepidermal moisture detection of pressure induced tissue damage on the trunk: The pressure ulcer detection study outcomes.

Authors:  Barbara M Bates-Jensen; Heather E McCreath; Anabel Patlan
Journal:  Wound Repair Regen       Date:  2017-05-31       Impact factor: 3.617

3.  Subepidermal moisture detection of heel pressure injury: The pressure ulcer detection study outcomes.

Authors:  Barbara M Bates-Jensen; Heather E McCreath; Gojiro Nakagami; Anabel Patlan
Journal:  Int Wound J       Date:  2017-12-17       Impact factor: 3.315

4.  Optical tissue probing: human skin hydration detection by speckle patterns analysis.

Authors:  Yarden Tzabari Kelman; Sagie Asraf; Nisan Ozana; Nadav Shabairou; Zeev Zalevsky
Journal:  Biomed Opt Express       Date:  2019-08-28       Impact factor: 3.732

5.  Characterizing stratum corneum structure, barrier function, and chemical content of human skin with coherent Raman scattering imaging.

Authors:  Sam Osseiran; Jomer Dela Cruz; Sinyoung Jeong; Hequn Wang; Christina Fthenakis; Conor L Evans
Journal:  Biomed Opt Express       Date:  2018-11-26       Impact factor: 3.732

6.  Subepidermal moisture is associated with early pressure ulcer damage in nursing home residents with dark skin tones: pilot findings.

Authors:  Barbara M Bates-Jensen; Heather E McCreath; Voranan Pongquan
Journal:  J Wound Ostomy Continence Nurs       Date:  2009 May-Jun       Impact factor: 1.741

7.  Abnormal expression of interleukin-6 is associated with epidermal alternations in localized scleroderma.

Authors:  Xingyu Zhu; Liuyiqi Jiang; Qian Zhong; Xiangzhen Kong; Rui Zhang; Lingxian Zhu; Qingmei Liu; Wenyu Wu; Yimei Tan; Jiucun Wang; Jingjing Xia
Journal:  Clin Rheumatol       Date:  2022-03-12       Impact factor: 2.980

8.  Formation and stability of oil-in-water nanoemulsions containing rice bran oil: in vitro and in vivo assessments.

Authors:  Daniela S Bernardi; Tatiana A Pereira; Naira R Maciel; Josiane Bortoloto; Gisely S Viera; Gustavo C Oliveira; Pedro A Rocha-Filho
Journal:  J Nanobiotechnology       Date:  2011-09-28       Impact factor: 10.435

9.  Efficacy of korean red ginseng in the treatment of atopic dermatitis.

Authors:  Kyung Goo Lee; Sang Wook Son
Journal:  J Ginseng Res       Date:  2011-06       Impact factor: 6.060

10.  Preparation, characterization and evaluation of sun protective and moisturizing effects of nanoliposomes containing safranal.

Authors:  Shiva Golmohammadzadeh; Fatemeh Imani; Hossein Hosseinzadeh; Mahmoud Reza Jaafari
Journal:  Iran J Basic Med Sci       Date:  2011-11       Impact factor: 2.699

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.