Literature DB >> 14729528

Development of gold stents for the treatment of intracranial aneurysms: an experimental study in a canine model.

Takashiro Ohyama1, Takuji Nishide, Hiroo Iwata, Waro Taki.   

Abstract

BACKGROUND AND
PURPOSE: Gold has often been used in medicine because of its radiopacity and flexibility. To perform stent-supported coil embolization of intracranial aneurysms, we prepared a gold stent and examined its flexibility, radiopacity, and thrombogenic properties in comparison with a stainless steel device implanted in vitro and in vivo.
METHODS: Gold stents were prepared by plating gold on stainless steel stents as a template. Their mechanical properties and trackability in vitro were determined and compared with those of stainless steel stents of the same design. Twenty gold stents and two stainless steel stents were implanted in canine external carotid, vertebral, and renal arteries, as a muscle branch of the maxillary arteries, to examine their performance in vivo.
RESULTS: The gold stent exhibited much less radial force and greater flexibility than the stainless steel stent. It also demonstrated superior trackability and radiopacity in the experimental endovascular procedures in canines. Histologic examination showed good patency of the stented artery with slight endothelial hypertrophy.
CONCLUSION: Although there is still room for more radial strength, less influence on intimal hypertrophy, a more suitable flexibility, and a smoother surface, the superior trackability and radiopacity of gold stents seem to support use of this device for the endovascular treatment of intracranial aneurysms.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14729528      PMCID: PMC7974184     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  25 in total

1.  Increased risk of restenosis after placement of gold-coated stents: results of a randomized trial comparing gold-coated with uncoated steel stents in patients with coronary artery disease.

Authors:  A Kastrati; A Schömig; J Dirschinger; J Mehilli; N von Welser; J Pache; H Schühlen; T Schilling; C Schmitt; F J Neumann
Journal:  Circulation       Date:  2000-05-30       Impact factor: 29.690

2.  Encrustation of intraprostatic stents--a comparative study.

Authors:  S A Holmes; P D Miller; P R Crocker; R S Kirby
Journal:  Br J Urol       Date:  1992-04

3.  Reaction of the aortic wall to six metallic stent materials.

Authors:  N Tanigawa; S Sawada; M Kobayashi
Journal:  Acad Radiol       Date:  1995-05       Impact factor: 3.173

4.  The "Remodelling Technique" in the Treatment of Wide Neck Intracranial Aneurysms. Angiographic Results and Clinical Follow-up in 56 Cases.

Authors:  J Moret; C Cognard; A Weill; L Castaings; A Rey
Journal:  Interv Neuroradiol       Date:  2001-05-15       Impact factor: 1.610

5.  Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity.

Authors:  B B Lieber; A P Stancampiano; A K Wakhloo
Journal:  Ann Biomed Eng       Date:  1997 May-Jun       Impact factor: 3.934

Review 6.  Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery. Case report and review of the literature.

Authors:  R T Higashida; W Smith; D Gress; R Urwin; C F Dowd; P A Balousek; V V Halbach
Journal:  J Neurosurg       Date:  1997-12       Impact factor: 5.115

7.  The electrochemical behavior of metallic implant materials as an indicator of their biocompatibility.

Authors:  H Zitter; H Plenk
Journal:  J Biomed Mater Res       Date:  1987-07

8.  Self-expanding and balloon-expandable stents in the treatment of carotid aneurysms: an experimental study in a canine model.

Authors:  A K Wakhloo; F Schellhammer; J de Vries; J Haberstroh; M Schumacher
Journal:  AJNR Am J Neuroradiol       Date:  1994-03       Impact factor: 3.825

9.  Embolization of experimentally created aneurysms with intravascular stent devices.

Authors:  G Geremia; M Haklin; L Brennecke
Journal:  AJNR Am J Neuroradiol       Date:  1994-08       Impact factor: 3.825

10.  Hydrophilic surface modification of metallic endoluminal stents.

Authors:  J M Seeger; M D Ingegno; E Bigatan; N Klingman; D Amery; C Widenhouse; E P Goldberg
Journal:  J Vasc Surg       Date:  1995-09       Impact factor: 4.268

View more
  3 in total

1.  The use of balloon-expandable stents in the management of intracranial arterial diseases: a 5-year single-center experience.

Authors:  Iruena Moraes Kessler; Charbel Mounayer; Michel Piotin; Laurent Spelle; Jose Ricardo Vanzin; Jacques Moret
Journal:  AJNR Am J Neuroradiol       Date:  2005-10       Impact factor: 3.825

2.  Goldvalve detachable balloon: "in vitro" assessment of safety and imaging artifacts in a 3-T MR system.

Authors:  Guilherme Dabus; Ronald J Gerstle; James Goodman; DeWitte T Cross; Christopher J Moran; Colin P Derdeyn
Journal:  Neuroradiology       Date:  2006-04-26       Impact factor: 2.804

3.  Evaluation of a Zn-2Ag-1.8Au-0.2V Alloy for Absorbable Biocompatible Materials.

Authors:  Ping Li; Christine Schille; Ernst Schweizer; Evi Kimmerle-Müller; Frank Rupp; Xingting Han; Alexander Heiss; Andreas Richter; Claudia Legner; Ulrich E Klotz; Jürgen Geis-Gerstorfer; Lutz Scheideler
Journal:  Materials (Basel)       Date:  2019-12-20       Impact factor: 3.623

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.