Literature DB >> 14727754

The use of a Leksell-BRW adapter for linac radiosurgery as an adjunct to Gamma Knife treatment.

Kenneth E Ekstrand1, William H Hinson, J Daniel Bourland, Allan F deGuzman, Volker W Stieber, Steven B Tatter, Thomas L Ellis.   

Abstract

We have investigated the use of an adapter that permits the use of a Leksell coordinate frame with a linear accelerator stereotactic radiosurgery system based on the Brown-Robert-Wells (BRW) design. This device is useful when lesions that are planned for treatment on a Leksell Gamma Knife system are found to be inaccessible to the Gamma Knife. We have found that with this device objects within a head phantom can be targeted by the linear accelerator within an accuracy of approximately 1 mm.

Mesh:

Year:  2003        PMID: 14727754     DOI: 10.1088/0031-9155/48/24/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Simulated gamma knife head frame placement for radiosurgical pre-planning.

Authors:  D Wiant; J D Bourland
Journal:  Technol Cancer Res Treat       Date:  2009-08

Review 2.  Current concepts in stereotactic radiosurgery - a neurosurgical and radiooncological point of view.

Authors:  Jan Vesper; B Bölke; C Wille; P A Gerber; C Matuschek; M Peiper; H J Steiger; W Budach; G Lammering
Journal:  Eur J Med Res       Date:  2009-03-17       Impact factor: 2.175

  2 in total