Bruno Manzo1, Giuseppe Liistro, Hugo De Clerck.
Abstract
The purpose of this clinical trial was to evaluate the reliability and time saved with a plasma arc curing unit (Apollo 95E, Dental/Medical Diagnostic Systems, Woodland Hills, Calif) compared with a conventional curing unit (Ortholux XL 3000, 3M Unitek, St Paul, Minn) for direct bracket bonding with resin adhesive. Forty-five patients were involved in the study, and 608 brackets were bonded in a contralateral quadrant pattern. The patients were followed for a mean (+/- standard deviation) period of 11 +/- 3.2 months. Survival analysis was carried out to compare the failure rate for the 2 techniques. The time required for bonding with each technique was also recorded. The mean survival time was 399 days, and there were no significant differences in survival time between the 2 bonding methods. Twelve bonding failures were reported with each technique. The curing time per bracket was significantly reduced with the plasma curing light compared with a conventional curing unit (65 +/- 19 vs 82 +/- 31 seconds). The plasma arc curing light can save chair-time without affecting the bonding failure rate.Entities:
Mesh:
Substances:
Year: 2004 PMID: 14718876 DOI: 10.1016/j.ajodo.2003.03.003
Source DB: PubMed Journal: Am J Orthod Dentofacial Orthop ISSN: 0889-5406 Impact factor: 2.650