Literature DB >> 14718829

Disappointing results with a new commercially available thoracic endograft.

Germano Melissano1, Yamume Tshomba, Efrem Civilini, Roberto Chiesa.   

Abstract

INTRODUCTION: New devices for endovascular treatment of thoracic aortic diseases were recently approved for clinical use by European authorities, obtaining the Conformité Européenne (CE) mark. In all patients who underwent endovascular treatment of a thoracic aortic disease in 2002, we used a new CE-marked device, the Endofit stent graft. The device is constituted of nitinol stents and polytetrafluoroethylene fabric, and has a simple design and delivery system.
METHODS: During 2002, 11 patients (mean age, 75 years; range, 66-85 years) underwent treatment of atherosclerotic aneurysm (n = 9), chronic type B dissection (n = 1), and intramural hematoma (n = 1). Disease involved the descending thoracic aorta in 7 patients and the distal aortic arch in 4 patients.
RESULTS: In all cases the Endofit stent grafts were successfully deployed in the intended position. No postoperative paraplegia or paraparesis was recorded. There were two in-hospital deaths: 1 patient died in the operating room (postmortem examination showed a kinked graft); and the other patient died in the intensive care unit on postoperative day 30, after an intraoperative stroke. One surgical conversion was performed 2 weeks after the procedure, because of total collapse of the graft due to rupture of three stents. Other graft-related complications included type I endoleak (n = 2), type II endoleak (n = 1), and incomplete opening of the device (n = 1).
CONCLUSION: Endovascular treatment of thoracic disease with the Endofit graft in this small heterogeneous group of patients resulted in several complications, which may arise from both the delivery system and the graft itself. At present, other commercially available endografts may be safer for endovascular treatment of thoracic aortic diseases.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14718829     DOI: 10.1016/s0741-5214(03)01034-6

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  5 in total

1.  CT appearance of complications related to thoracic endovascular aortic repair (TEVAR): a pictorial essay.

Authors:  U Pua; K H Tay; B S Tan; M M Htoo; M Sebastian; K Sin; Y L Chua
Journal:  Eur Radiol       Date:  2009-01-14       Impact factor: 5.315

2.  Hybrid endovascular repair in aortic arch pathologies: a retrospective study.

Authors:  Xiaohui Ma; Wei Guo; Xiaoping Liu; Tai Yin; Xin Jia; Jiang Xiong; Hongpeng Zhang; Lijun Wang
Journal:  Int J Mol Sci       Date:  2010-11-18       Impact factor: 5.923

3.  Delayed endovascular treatment of descending aorta stent graft collapse in a patient treated for post- traumatic aortic rupture: a case report.

Authors:  Giovanni Nano; Daniela Mazzaccaro; Giovanni Malacrida; Maria Teresa Occhiuto; Silvia Stegher; Domenico G Tealdi
Journal:  J Cardiothorac Surg       Date:  2011-05-24       Impact factor: 1.637

4.  Delayed Aortic Stent Collapse in Blunt Traumatic Aortic Injury Repair.

Authors:  Abdullah Alhaizaey; Badr Aljabri; Musaad Alghamdi; Ali AlAhmari; Ahmed Abulyazied; Mohammed Asiry; Mohammed Al-Omran
Journal:  Aorta (Stamford)       Date:  2020-02-04

5.  Efficacy of thermoplastic polyurethane and gelatin blended nanofibers covered stent graft in the porcine iliac artery.

Authors:  Dae Sung Ryu; Dong-Sung Won; Ji Won Kim; Yubeen Park; Song Hee Kim; Jeon Min Kang; Chu Hui Zeng; Dohyung Lim; Hyun Choi; Jung-Hoon Park
Journal:  Sci Rep       Date:  2022-10-03       Impact factor: 4.996