Correspondence of Tono-Pen intraocular pressure measurements performed at the central cornea and mid-peripheral cornea.

Michael Sullivan-Mee1, Fortsmith Pham.   

Abstract

BACKGROUND: As the awareness of the influence of central corneal thickness (CCT) on Goldmann tonometry has increased, many publications have questioned the accuracy of Goldmann intraocular pressure (IOP) measurement. The Tono-Pen, because it indents a much smaller surface area when compared to a Goldmann probe, may be less affected by corneal thickness variations when compared with Goldmann tonometry.
METHODS: Forty human subjects with no history of refractive surgery participated in this study. The IOP of the right eye of each subject was measured with the Goldmann tonometer, the Tono-Pen at the central cornea, and the Tono-Pen at the mid-peripheral cornea. An ultrasonic DGH Pachette pachymeter was used to measure the central and mid-peripheral corneal thickness at the location of the IOP readings.
RESULTS: Tono-Pen measurements at the central and mid-peripheral cornea highly correlated (r = 0.933), and did not significantly differ (p = 0.646). The IOP readings with the Goldmann tonometer (r= 0.406), the Tono-Pen at the central cornea (r = 0.453), and the Tono-Pen at the mid-peripheral cornea (r = 0.321) showed a positive correlation to corneal thickness. The Goldmann and Tono-Pen tonometers differed significantly in the measurement of IOP at the central cornea (p = 0.007), but were positively correlated (r = 0.674).
CONCLUSIONS: The Tono-Pen IOP measurement at the central cornea highly approximated Tono-Pen IOP measurement at the mid-peripheral cornea. Furthermore, although not highly correlated, both the Goldmann and Tono-Pen tonometers showed a significantly positive correlation between IOP and corneal thickness measurements.

Entities:  

Mesh:

Year:  2004        PMID: 14717278     DOI: 10.1016/s1529-1839(04)70008-1

Source DB:  PubMed          Journal:  Optometry        ISSN: 1558-1527


  2 in total

1.  Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.

Authors:  Mahmoud Rateb; Mahmoud Abdel-Radi; Zeiad Eldaly; Mohamed Nagy Elmohamady; Asaad Noor El Din
Journal:  J Ophthalmol       Date:  2019-05-09       Impact factor: 1.909

2.  Corneal biomechanical parameters and intraocular pressure: the effect of topical anesthesia.

Authors:  Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2012-06-15