| Literature DB >> 14717187 |
Ezra C Wood1, Paul J Wooldridge, Jens H Freese, Tim Albrecht, Ronald C Cohen.
Abstract
We describe a prototype designed for in situ detection of the nitrate radical (NO3) by laser-induced fluorescence (LIF) and of N2O5 by thermal dissociation followed by LIF detection of NO3. An inexpensive 36 mW continuous wave multi-mode diode laser at 662 nm is used to excite NO3 in the B2E'(0000) <-- X2A'2(0000) band. Fluorescence is collected from 700 to 750 nm. The prototype has a sensitivity to NO3 of 76 ppt for a 60 s integration with an accuracy of 8%. Although this sensitivity is adequate for studies of N205 in many environments, it is much less sensitive (about 300 times) than expected based on a comparison of previously measured photophysical properties of NO2 and NO3. This implies much stronger nonradiative coupling of electronic states in NO3 than in NO2.Entities:
Mesh:
Substances:
Year: 2003 PMID: 14717187 DOI: 10.1021/es034507w
Source DB: PubMed Journal: Environ Sci Technol ISSN: 0013-936X Impact factor: 9.028