Literature DB >> 14712950

Ultrasound assessment of bladder volume: is it valid after delivery?

Lucinda M Pallis1, Melissa Wilson.   

Abstract

AIM: To assess the validity of a commercially available bladder volume scanner in the puerperium.
METHODS: A prospective blinded comparison of bladder volume measurement after vaginal delivery using the BladderScan bladder volume instrument (BVI) 3000 and Foley catheter; comparison using the intraclass correlation coefficient.
RESULTS: The mean difference between the two measurements was a 130-mL over-measurement by the bladder scanner (range: -156 mL to +422 mL). The intraclass correlation coefficient was 0.23 (95% confidence interval 0.00, 0.59).
CONCLUSIONS: The BladderScan BVI 3000 is not an accurate instrument to assess bladder volume the day after vaginal delivery. Some of the discrepancy might relate to use of the Foley catheter as the reference standard. Further comparison between the BladderScan and a short female catheter or real time ultrasound is indicated.

Mesh:

Year:  2003        PMID: 14712950     DOI: 10.1046/j.0004-8666.2003.00141.x

Source DB:  PubMed          Journal:  Aust N Z J Obstet Gynaecol        ISSN: 0004-8666            Impact factor:   2.100


  5 in total

Review 1.  Urinary Retention in Pregnancy and Puerperium: Acupuncture Treatment.

Authors:  Poovadan Sudhakaran
Journal:  Med Acupunct       Date:  2019-10-17

2.  Validity of bladder volume measurement by ultrasound in women postpartum.

Authors:  Josefine Tangen Jensen; Niels Klarskov; Jeannet Lauenborg
Journal:  Int Urogynecol J       Date:  2019-07-05       Impact factor: 2.894

3.  Is Doppler planimetry a valid technique for the evaluation of postpartum urinary bladder volume?

Authors:  James W S Lee; Stergios K Doumouchtsis; Michelle M Fynes
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-01-30

4.  Accuracy of postvoid residual volumes after vaginal delivery: a prospective equivalence study to compare an automatic scanning device with transurethral catheterization.

Authors:  Femke E M Mulder; Sytske van der Velde; Fraukje Pol; Marjolein Bos; Jules Schagen van Leeuwen; Viviane Dietz; Robert A Hakvoort; Jan-Paul W R Roovers
Journal:  Int Urogynecol J       Date:  2018-06-27       Impact factor: 2.894

5.  Three-dimensional bladder ultrasonography with the BladderScan® overestimates post void residual one week after delivery.

Authors:  Simone Salimans; Jolien Govaerts; Natasja de Jong; Jeroen van Bavel; Leonie Speksnijder
Journal:  Eur J Obstet Gynecol Reprod Biol X       Date:  2019-05-14