Literature DB >> 14711708

Plastic particles at the LASIK interface.

Anders Ivarsen1, Jan Thøgersen, Søren Rud Keiding, Jesper Ø Hjortdal, Torben Møller-Pedersen.   

Abstract

PURPOSE: To investigate the origin, composition, and persistence of the interface particles that frequently are observed after LASIK.
DESIGN: Small case series and experimental animal study.
METHODS: Four patients received LASIK using a Schwind Supratome (Schwind, Kleinostheim, Germany) and a MEL 70 G-Scan excimer laser (Asclepion, Jena, Germany) and were examined over the course of 1 year using slit-lamp and in vivo confocal microscopy. Four rabbits received a monocular microkeratome incision and were examined immediately after surgery without lifting the flap. After monthly evaluation for 4 months using in vivo confocal microscopy, 2 corneas were processed for histologic analysis and were sectioned serially. To measure the iron content, atomic absorption spectrometry was performed on 2 operated and 2 unoperated rabbit corneas. The chemical composition of the metal and plastic parts of the microkeratome blade was identified using energy dispersive x-ray fluorescence (metal part), and Raman and infrared spectroscopy (plastic part). Before and after oscillation in air, the microkeratome blade and motor-head were examined using light and fluorescence microscopy. In serial sections, interface particles were identified by fluorescence microscopy and their chemical composition was determined using Coherent Antistokes Raman Scattering microscopy.
RESULTS: In LASIK patients, thousands of brightly reflecting particles (up to 30 micro m) were observed throughout the interface. The highest particle density was detected where the microkeratome blade had first entered the cornea. Both in the center and at the flap edge, the morphologic features, distribution, and density of these particles remained unaltered throughout the 1-year observation period. In rabbit corneas, interface particles were observed immediately after the microkeratome incision, even though the flap had not been lifted. These particles were similar to those observed in humans and persisted unaltered throughout the study. The operated and unoperated rabbit corneas had comparable iron content, demonstrating that the particles were not fragments of the uncoated steel blade. Only a few particles were observed on the unused microkeratome motor head and blade, whereas numerous fluorescent particles were detected after oscillation in air, the amount of particles increasing with oscillation time. Interestingly, the only fluorescent part of the microkeratome was the plastic segment of the blade. This plastic (polyetherimide) emitted fluorescence identical to that of the observed particles, whereas all metal parts of the microkeratome blade and motor head were nonfluorescent. In serial sections, interface particles showed fluorescent properties equivalent to polyetherimide and exhibited molecular resonance at 1780 and 3100 cm(-1), in accordance with the Raman spectrum of polyetherimide.
CONCLUSIONS: Numerous plastic particles are generated during microkeratome oscillation and are deposited at the interface during LASIK. The particles persist unaltered for at least 1 year.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14711708     DOI: 10.1016/j.ophtha.2003.05.021

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  4 in total

1.  Corneal response to femtosecond laser photodisruption in the rabbit.

Authors:  Naoyuki Morishige; Anna Kesler-Diaz; Andrew J Wahlert; Ronald M Kurtz; Tibor Juhasz; Melvin Sarayba; James V Jester
Journal:  Exp Eye Res       Date:  2008-03-06       Impact factor: 3.467

2.  In vivo biopsy by laser confocal microscopy for evaluation of traumatic recurrent corneal erosion.

Authors:  Tai-ichiro Chikama; Norihisa Takahashi; Makiko Wakuta; Naoyuki Morishige; Teruo Nishida
Journal:  Mol Vis       Date:  2008-12-15       Impact factor: 2.367

3.  In vivo evaluation of DSAEK interface with scanning-laser confocal microscopy.

Authors:  Giulio Ferrari; Verena Reichegger; Luca Ludergnani; Elisabetta Delfini; Claudio Macaluso
Journal:  BMC Ophthalmol       Date:  2012-08-01       Impact factor: 2.209

4.  Spontaneous, uncomplicated dissolution of a large cotton fiber in the laser in situ keratomileusis interface.

Authors:  Gaurav Prakash; Namrata Sharma; Vandana Choudary; Jeewan S Titiyal
Journal:  Middle East Afr J Ophthalmol       Date:  2012 Jul-Sep
  4 in total