Calibration of ion effective temperatures achieved by resonant activation in a quadrupole ion trap.

Valérie Gabelica1, Michael Karas, Edwin De Pauw.   

Abstract

The present paper describes a calibration of the ion effective temperatures as a function of the resonant activation amplitude in a quadrupole ion trap mass spectrometer. MS/MS experiments are performed on leucine enkephalin (M + H)+, bradykinin (M + H)+, (M + 2H)2+, and (M + 3H)3+, and ubiquitin (M + 11H)11+. For each amplitude, the effective temperature is calculated as the temperature that would give the same dissociation rate constant as the one observed and is calculated using published Arrhenius parameters. The effective temperature is found to be linearly dependent on the activation amplitude on the range investigated. The dependence of the slope and of the intercept of the T(eff) = f (amplitude) functions on the parent ion m/z is examined and an equation is derived to calibrate the ion effective temperature between 365 and 600 K. Below 365 K, a deviation from linearity is expected. Above 600 K, the validity of the equation will depend on whether the rapid energy exchange limit is still reached. Calculating backward, the Arrhenius parameters from the measured dissociation rates using this calibration show excellent agreement with the published values. The calibration can therefore be used to determine Arrhenius activation parameters from dissociation kinetics under resonant activation in quadrupole ion trap mass spectrometers.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14708790     DOI: 10.1021/ac034473a

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  3 in total

1.  In silico identification software (ISIS): a machine learning approach to tandem mass spectral identification of lipids.

Authors:  Lars J Kangas; Thomas O Metz; Giorgis Isaac; Brian T Schrom; Bojana Ginovska-Pangovska; Luning Wang; Li Tan; Robert R Lewis; John H Miller
Journal:  Bioinformatics       Date:  2012-05-15       Impact factor: 6.937

2.  Collisional activation of ions in RF ion traps and ion guides: the effective ion temperature treatment.

Authors:  Aleksey V Tolmachev; Andrey N Vilkov; Bogdan Bogdanov; Ljiljana Pasa-Tolić; Christophe D Masselon; Richard D Smith
Journal:  J Am Soc Mass Spectrom       Date:  2004-11       Impact factor: 3.109

3.  Modular calibrant sets for the structural analysis of nucleic acids by ion mobility spectrometry mass spectrometry.

Authors:  Jennifer L Lippens; Srivathsan V Ranganathan; Rebecca J D'Esposito; Daniele Fabris
Journal:  Analyst       Date:  2016-06-20       Impact factor: 4.616

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.