Study on the tongue and groove effect of the Elekta multileaf collimator using Monte Carlo simulation and film dosimetry.

Freddy Haryanto1, Matthias Fippel, Annemarie Bakai, Fridtjof Nüsslin.   

Abstract

BACKGROUND: Nowadays, multileaf collimation of the treatment fields from medical linear accelerators is a common option. Due to the design of the leaf sides, the tongue and groove effect occurs for certain multileaf collimator applications such as the abutment of fields where the beam edges are defined by the sides of the leaves.
MATERIAL AND METHODS: In this study, the tongue and groove effect was measured for two pairs of irregular multileaf collimator fields that were matched along leaf sides in two steps. Measurements were made at 10 cm depth in a polystyrene phantom using Kodak EDR2 films for a photon beam energy of 6 MV on an Elekta Sli-plus accelerator. To verify the measurements, full Monte Carlo simulations were done. In the simulations, the design of the leaf sides was taken into account and one component module of BEAM code was modified to correctly simulate the Elekta multileaf collimator. RESULTS AND
CONCLUSION: The results of measurements and simulations are in good agreement and within the tolerance of film dosimetry.

Entities:  

Mesh:

Year:  2004        PMID: 14704846     DOI: 10.1007/s00066-004-1135-3

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  4 in total

1.  CT-myelography for high-dose irradiation of spinal and paraspinal tumors with helical tomotherapy: revival of an old tool.

Authors:  Matthias Uhl; Florian Sterzing; Gregor Habl; Kai Schubert; Gabriele Sroka-Perez; Jürgen Debus; Klaus Herfarth
Journal:  Strahlenther Onkol       Date:  2011-06-27       Impact factor: 3.621

2.  Evaluating the peak-to-valley dose ratio of synchrotron microbeams using PRESAGE fluorescence.

Authors:  N Annabell; N Yagi; K Umetani; C Wong; M Geso
Journal:  J Synchrotron Radiat       Date:  2012-03-15       Impact factor: 2.616

3.  Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems.

Authors:  Sanaz Hariri; Majid Shahriari
Journal:  J Appl Clin Med Phys       Date:  2010-06-15       Impact factor: 2.102

4.  Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams.

Authors:  R Mohan; K Jayesh; R C Joshi; Maha Al-Idrisi; P Narayanamurthy; Saroj Kumar Das Majumdar
Journal:  J Med Phys       Date:  2008-07
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.