Eccentricity and measurement variability and repeatability with the retinal thickness analyser.

E D Gilmore1, C Hudson.   

Abstract

AIMS: To define the variability and repeatability of retinal thickness measurements using the retinal thickness analyser (RTA) and to elucidate any interaction between eccentricity (that is, position relative to the fovea) and variability and repeatability.
METHODS: The sample comprised 20 normal subjects of mean age 33 years. Each subject attended for two visits. Repeated RTA scans were acquired centred on the fovea and for any one of the four possible non-foveal scan areas. The mean retinal thickness (+SD) was calculated for a series of concentric circular bands centred on fixation. A repeated measures analysis of variance (ANOVA) was used to determine any significant interaction between the variability of RTA thickness values and eccentricity.
RESULTS: The group mean coefficient of variation and coefficient of repeatability were highest at the fovea. The repeated measures ANOVA revealed that the within test variability of RTA measurements varied significantly with eccentricity (p<0.0001). Similarly, the between test repeatability varied significantly with eccentricity (p = 0.045).
CONCLUSION: The significantly elevated within test variability and between test repeatability in the foveal area need to be considered when using the RTA to evaluate patients with macular disease.

Entities:  

Mesh:

Year:  2004        PMID: 14693775      PMCID: PMC1771935          DOI: 10.1136/bjo.88.1.62

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  35 in total

1.  Application of rapid scanning retinal thickness analysis in retinal diseases.

Authors:  S Asrani; R Zeimer; M F Goldberg; S Zou
Journal:  Ophthalmology       Date:  1997-07       Impact factor: 12.079

2.  A new method for rapid mapping of the retinal thickness at the posterior pole.

Authors:  R Zeimer; M Shahidi; M Mori; S Zou; S Asrani
Journal:  Invest Ophthalmol Vis Sci       Date:  1996-09       Impact factor: 4.799

3.  Foveal cysts: a premacular hole condition associated with vitreous traction.

Authors:  J C Folk; H C Boldt; D G Keenum
Journal:  Arch Ophthalmol       Date:  1998-09

4.  Quantitative in vivo retinal thickness measurements in healthy subjects.

Authors:  D Landau; E M Schneidman; T Jacobovitz; Y Rozenman
Journal:  Ophthalmology       Date:  1997-04       Impact factor: 12.079

5.  Quantitative analysis of foveal retinal thickness in diabetic retinopathy with the scanning retinal thickness analyzer.

Authors:  T Yasukawa; J Kiryu; A Tsujikawa; J Dong; I Suzuma; H Takagi; Y Ogura
Journal:  Retina       Date:  1998       Impact factor: 4.256

6.  Topography of the adult human retina.

Authors:  B R Straatsma; M B Landers; A E Kreiger; L Apt
Journal:  UCLA Forum Med Sci       Date:  1969

7.  Clinical assessment of the macula by retinal topography and thickness mapping.

Authors:  J P Gieser; M M Rusin; M Mori; N P Blair; M Shahidi
Journal:  Am J Ophthalmol       Date:  1997-11       Impact factor: 5.258

8.  Quantitative detection of glaucomatous damage at the posterior pole by retinal thickness mapping. A pilot study.

Authors:  R Zeimer; S Asrani; S Zou; H Quigley; H Jampel
Journal:  Ophthalmology       Date:  1998-02       Impact factor: 12.079

9.  Serial optical sectioning of macular holes at different stages of development.

Authors:  S Asrani; R Zeimer; M F Goldberg; S Zou
Journal:  Ophthalmology       Date:  1998-01       Impact factor: 12.079

10.  Retinal thickness change after focal laser treatment of diabetic macular oedema.

Authors:  M Shahidi; Y Ogura; N P Blair; R Zeimer
Journal:  Br J Ophthalmol       Date:  1994-11       Impact factor: 4.638

View more
  2 in total

Review 1.  A reference standard for the measurement of macular oedema.

Authors:  K A Goatman
Journal:  Br J Ophthalmol       Date:  2006-09       Impact factor: 4.638

2.  Repeatability of Foveal Measurements Using Spectralis Optical Coherence Tomography Segmentation Software.

Authors:  Irene Ctori; Byki Huntjens
Journal:  PLoS One       Date:  2015-06-15       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.