Literature DB >> 14678813

Can causes of false-normal D-dimer test [SimpliRED] results be identified?

R A Kraaijenhagen1, J Wallis, M M W Koopman, M R de Groot, F Piovella, P Prandoni, H R Büller.   

Abstract

BACKGROUND: To simplify the diagnostic strategy of patients suspected for venous thromboembolism, the use of D-dimer tests has been advocated. Very important for the safety of such diagnostic strategies would be the capacity to recognise false-normal D-dimer results, in order to prevent inadequately withholding anticoagulant treatment in patients who actually have the disease. Insight in the causes of false-normal D-dimer results would therefore be necessary. We hypothesised that certain patient characteristics are associated with relatively low plasma D-dimer levels and, therefore, could increase the risk of false-normal results.
METHODS: Consecutive patients with an objectively confirmed venous thromboembolic event and an independently obtained false-normal SimpliRED D-dimer test result were included in the study. For each patient, two controls with objectively confirmed venous thromboembolism and an adequate abnormal D-dimer result were selected. Baseline patient characteristics, obtained by standardised questionnaires, were compared between the two groups of patients.
RESULTS: In total, 686 patients had a venous thromboembolic event and 47 of these patients had a false-normal SimpliRED result. Therefore, the overall sensitivity of the SimpliRED test for venous thromboembolism was 94% (95% CI: 92-95%). Although the prevalence of certain clinical characteristics was significantly higher in patients with a false-normal D-dimer result than in the controls [odds ratios for (LMW)heparin treatment and symptoms lasting more than 10 days: 5.1 (95% CI: 1.5-18.7) and 3.2 (95% CI:1.4-7.4), respectively], the prevalence of these characteristics was also high in the control group with an adequate abnormal D-dimer. Combining two or more of these characteristics had a low prevalence and did not further improve the ability to identify those patients with a false-normal D-dimer test at presentation.
CONCLUSIONS: Although these findings clearly indicate an association between certain baseline clinical characteristics and the occurrence of a false-normal SimpliRED test, the clinical utility for these characteristics is limited.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14678813     DOI: 10.1016/j.thromres.2003.08.028

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  4 in total

1.  The great deception: tranexamic acid and extensive pulmonary emboli.

Authors:  Amr Salam; Christopher King; Orhan Orhan; Vincent Mak
Journal:  BMJ Case Rep       Date:  2013-01-31

2.  Evidence of Disseminated Intravascular Coagulation in a Porcine Model Following Radiation Exposure.

Authors:  G S Krigsfeld; J B Shah; J K Sanzari; L Lin; A R Kennedy
Journal:  Life Sci Space Res (Amst)       Date:  2014-10-01

Review 3.  Excluding venous thromboembolism using point of care D-dimer tests in outpatients: a diagnostic meta-analysis.

Authors:  G J Geersing; K J M Janssen; R Oudega; L Bax; A W Hoes; J B Reitsma; K G M Moons
Journal:  BMJ       Date:  2009-08-14

4.  Identification of and solution for false D-dimer results.

Authors:  Xian-Yan Zhang; Xue-Xuan Zhang; Jia-Long Xu; Teng-Yi Huang; Ying Wu; Ye-Ru Yang; Huan-Bin Zhou; Ying-E Wu
Journal:  J Clin Lab Anal       Date:  2020-01-22       Impact factor: 2.352

  4 in total