| Literature DB >> 14677129 |
Abstract
Calculated and measured resolution figures are presented for confocal microscopes with different pinhole sizes and for nonlinear (2-photon and second harmonic) microscopes. A modest degree of super-resolution is predicted for a confocal microscope but in practice this is not achievable and confocal fluorescence gives little resolution improvement over widefield. However, practical non-linear microscopes do approach their theoretical resolution and therefore show no resolution disadvantage relative to confocal microscopes in spite of the longer excitation wavelength. Copyright 2003 Wiley-Liss, Inc.Mesh:
Substances:
Year: 2004 PMID: 14677129 DOI: 10.1002/jemt.10423
Source DB: PubMed Journal: Microsc Res Tech ISSN: 1059-910X Impact factor: 2.769