Literature DB >> 14676553

The magnetic attraction of lower extremity external fixators in an MRI suite.

Brian L Davison1, Robert V Cantu, Sue Van Woerkom.   

Abstract

OBJECTIVES: To determine the magnetic attraction of 10 currently available lower extremity external fixators in the MRI suite and to test fixators with minimal or no magnetic attraction for the presence of temperature elevation during a 30-minute MRI scan.
DESIGN: Consecutive testing of selected tibial external fixation models in an identical manner.
SETTING: Level 1 trauma center. CONSTRUCTS: Ten commercially available external fixators used for lower extremity injuries applied to saw bone tibia. INTERVENTION: External fixators suspended in testing apparatus were placed at three locations in the MRI suite for magnetic testing. Fixators with minimal attraction were scanned for 30 minutes. MAIN OUTCOME MEASUREMENTS: Magnetic attraction was measured with a digital scale, and temperature measurements were obtained with a digital thermometer and thermocouple.
RESULTS: The Ilizarov with carbon fiber rings, Richards Hex-Fix, and Large Synthes External Fixator had <1 kg of magnetic attraction at all three locations. No component of these fixators experienced >2 degrees of temperature elevation after a 30-minute MRI scan. The EBI Dynafix with Ankle Clamp, EBI Dynafix, and EBI Dynafix Hybrid along with the Hoffman II, Hoffman II Hybrid, Ilizarov with stainless steel rings, and Synthes Hybrid had >1 kg of magnetic attraction at all three locations.
CONCLUSIONS: Many commercially available external fixators have components and spring clamps that have significant magnetic attraction to the MRI scanner. Fixators that have <1 kg of attraction do not experience significant heating during MRI.

Entities:  

Mesh:

Year:  2004        PMID: 14676553     DOI: 10.1097/00005131-200401000-00005

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  MRI evaluation of the knee with non-ferromagnetic external fixators: cadaveric knee model.

Authors:  Peter Elsissy; Yusuf T Akpolat; Alexander Chien; Wayne K Cheng
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-06-09

2.  Safety and image quality of MR-conditional external fixators for 1.5 Tesla extremity MR.

Authors:  David H Ballard; John D Garrett; Alberto A Simoncini; Silvia Barbeito; Massimo Max Morandi
Journal:  Emerg Radiol       Date:  2021-01-15

3.  Magnetic Resonance Imaging of the Knee in the Presence of Bridging External Fixation: A Comparative Experimental Evaluation of Four External Fixators, Including Dolphix®.

Authors:  Francesco Addevico; Alberto Simoncini; Giovanni Solitro; Massimo Max Morandi
Journal:  J Funct Morphol Kinesiol       Date:  2021-12-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.