A method for high-precision and high-accuracy mass spectrometric measurements of the ratios among the three oxygen isotopes, and of the O(2)/Ar ratio, is presented. It involves separation of the O(2)-Ar mixture from air and includes a fully automated system that ensures highly reliable sample processing. Repeated measurements of atmospheric oxygen yield the repeatability (+/-SE x t, standard error of the mean (n = 12) multiplied by Student's t-factor for a 95% confidence limit) of 0.004, 0.003 and 0.2 per thousand for delta(18)O, delta(17)O and delta O(2)/Ar, respectively. Copyright 2003 John Wiley & Sons, Ltd.Entities: