Bacterial retention properties of heat and moisture exchange filters.

M G Lee1, J L Ford, P B Hunt, D S Ireland, P W Swanson.   

Abstract

We have examined the properties of six heat and moisture exchange filters (HMEF) to ascertain their resistance to liquid flow and their ability to retain a challenge bacterium, Pseudomonas diminuta, from aqueous and nebulized suspensions. Only one HMEF, the Pall Ultipor was able to withstand a significantly greater pressure of liquid than that found in clinical practice. However, when breached, the HMEF were unable to prevent transmission of micro-organisms from aqueous suspension. Only the Darex Hydrobac filter failed to meet the manufacturer's claim for filter efficiency for nebulized bacteria, mainly because the filter housing failed under test. When the reduction in bacterial cells after passage of the nebulized Pseudomonas diminuta through the HMEFs was analysed statistically, the data showed that the HMEF produced by Pall (Ultipor) and Intersurgical (Filter therm) were superior to those produced by DAR Mediplan (Hygrobac), Intertech (HME 225-2835-800) and Gibeck (Humid-vent).

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1467087     DOI: 10.1093/bja/69.5.522

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  1 in total

Review 1.  Humidification and heating of inhaled gas in patients with artificial airway. A narrative review.

Authors:  Gustavo Adrián Plotnikow; Matias Accoce; Emiliano Navarro; Norberto Tiribelli
Journal:  Rev Bras Ter Intensiva       Date:  2018-03