Evgenij V Potapov1, Yuguo Weng, Harald Hausmann, Michael Kopitz, Miralem Pasic, Roland Hetzer.
Abstract
In 12 patients with acute cardiogenic shock who required mechanical circulatory support a short-term Abiomed BVS 5000 extracorporeal assist device was implanted using the inflow and outflow cannulas of the BerlinHeart extracorporeal assist device. In 7 patients suitable for long-term support the Abiomed pumps were later exchanged for BerlinHeart pumps. This approach avoids the risks associated with repeat sternotomy and use of cardiopulmonary bypass and decreases the total costs of patient care.Entities:
Mesh:
Year: 2003 PMID: 14667665 DOI: 10.1016/s0003-4975(03)00693-3
Source DB: PubMed Journal: Ann Thorac Surg ISSN: 0003-4975 Impact factor: 4.330