Non invasive hemodynamic monitorization for AV interval optimization in patients with ventricular resynchronization therapy.

José Ferreira Santos1, Leonor Parreira, João Madeira, Nuno Fonseca, Luís Neves Soares, Lopes Inês.   

Abstract

INTRODUCTION AND
OBJECTIVE: In patients (pts) with dilated cardiomyopathy and intraventricular conduction disturbances, resynchronization therapy improves cardiac function and functional capacity. Determination of the optimal AV interval is essential to optimize the therapy results and various methods have been used in daily practice to obtain such an interval. The aim of this work is to assess optimal AV determined by impedance cardiography and compare it to the interval previously obtained by transmitral flow Doppler echocardiography. PATIENTS: Seven pts were evaluated, five men, mean age of 61 +/- 10 years, with dilated cardiomyopathy, intra-ventricular conduction disturbances and heart failure, New York Heart Association functional class III or IV. Four pts had ischemic cardiomyopathy and three idiopathic. All pts had an implanted cardioverter-defibrillator with cardiac resynchronization.
METHODS: The optimal AV delay was estimated by transmitral flow pulsed Doppler using the method previously described by Ritter. Subsequently, cardiac output (CO) was determined by impedance cardiography. CO was estimated for different AV delays, beginning with 80 ms until occurrence of fusion QRS. The optimal AV delay was defined as the value corresponding to the highest measured CO. The ideal AV interval was compared between the two methods used.
RESULTS: The AV delay determined by echocardiography varied between 120 and 170 ms (134 +/- 17 ms). The optimum AV delay based on CO estimation varied between 110 and 190 ms (137 +/- 26 ms). There was a correlation between the delays determined by the two methods (r = 0.844; p = 0.017).
CONCLUSIONS: In pts undergoing ventricular resynchronization therapy, AV delay optimization based on CO determined by impedance cardiography is comparable to that measured by transmitral flow pulsed Doppler. However, impedance cardiography seems a more objective and simpler technique.

Entities:  

Mesh:

Year:  2003        PMID: 14655311

Source DB:  PubMed          Journal:  Rev Port Cardiol        ISSN: 0870-2551            Impact factor:   1.374


  4 in total

1.  Comparison of the acute hemodynamic effect of right ventricular apex, outflow tract, and dual-site right ventricular pacing.

Authors:  Andrzej Rubaj; Piotr Rucinski; Tomasz Sodolski; Andrzej Bilan; Marcin Gulaj; Alicja Dabrowska-Kugacka; Andrzej Kutarski
Journal:  Ann Noninvasive Electrocardiol       Date:  2010-10       Impact factor: 1.468

2.  Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy.

Authors:  Martin U Braun; Andreas Schnabel; Thomas Rauwolf; Matthias Schulze; Ruth H Strasser
Journal:  J Interv Card Electrophysiol       Date:  2005-09       Impact factor: 1.900

3.  Novel insights on effect of atrioventricular programming of biventricular pacemaker in heart failure--a case series.

Authors:  Tasneem Z Naqvi; Asim M Rafique
Journal:  Cardiovasc Ultrasound       Date:  2006-10-16       Impact factor: 2.062

4.  Echocardiographic demonstration of the effect of varying paced a-v intervals on ventricular filling pattern.

Authors:  Damodar Kumbala; Sony Jacob; Masoor Kamalesh; Mithilesh Das
Journal:  Indian Pacing Electrophysiol J       Date:  2008-08-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.