| Literature DB >> 14655228 |
Andreas E Biedler1, Wolfram Wilhelm, Sascha Kreuer, Stefan Soltesz, Friedhelm Bach, Friedrich O Mertzlufft, Gerd P Molter.
Abstract
This study was designed to assess the pCO(2) accuracy of portable mainstream (Tidal Wave, Novametrix; Propaq 106, Protocol) and sidestream capnometers (Capnocheck 8200, BCI; Capnocount mini, Weinmann; NPB-75, Nellcor Puritan Bennett; SC-210, Pryon) with respect to international standards and preclinical emergency conditions. Measurements were performed under temperature conditions of +22 degrees C and -20 degrees C using dry gas mixtures with different CO(2) concentrations (STPD) and in patients ventilated with pure oxygen (BTPS). Accuracy presented to be between +1% (Capnocheck) and +12% (Propaq) (STPD) and between -0.4% (Capnocheck) and +11% (Tidal Wave) (BTPS). The measurements were affected by low ambient temperature only in the NPB-75 (+15%). Our results indicate that portable quantitative capnometers are able to fulfill accuracy requirements as requested by international standards but can be affected by changing ambient temperatures.Entities:
Mesh:
Year: 2003 PMID: 14655228 DOI: 10.1016/j.ajem.2003.08.005
Source DB: PubMed Journal: Am J Emerg Med ISSN: 0735-6757 Impact factor: 2.469