Membrane oxygenator exhaust capnography for continuously estimating arterial carbon dioxide tension during cardiopulmonary bypass.

Kieron C Potger1, Darryl McMillan, Joanne Southwell, Hayden Dando, Killian O'Shaughnessy.   

Abstract

Typically, the standard practice for measuring the arterial blood carbon dioxide tension (PaCO2) during cardiopulmonary bypass (CPB) is to take intermittent blood samples for analysis by a bench blood gas analyzer. Continuous inline blood gas monitors are available but are expensive. A potential solution is the capnograph, which was evaluated by determining how accurately the carbon dioxide tension in the oxygenator exhaust gases (PECO2) predicts PaCO2. A standard capnograph monitoring line was attached to the exhaust port of the membrane oxygenator. During CPB, the capnograph reading and arterial blood temperature were recorded at the same time as routine arterial blood gases were taken. One hundred fifty-seven blood samples were collected from 78 patients. A good correlation was found between the PECO2 and the temperature corrected PaCO2 (r2 = 0.833, P < .001). There was also a reasonable degree of agreement between the PECO2 and the temperature corrected PaCO2 during all phases of CPB: accuracy (bias or mean difference between PaCO2 and PECO2) of -1.2 mmHg; precision (95% limits of agreement) of +/- 4.7 mmHg. These results suggest that oxygenator exhaust capnography may be a simple and inexpensive adjunct to the bench blood gas analyzer in continuously estimating PaCO2 of a clinically useful degree of accuracy during CPB.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14653424

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  5 in total

1.  Oxygenator exhaust capnography for prediction of arterial carbon dioxide tension during hypothermic cardiopulmonary bypass.

Authors:  Anis Baraka; Mohamad El-Khatib; Eva Muallem; Salim Jamal; Sania Haroun-Bizri; Marie Aouad
Journal:  J Extra Corpor Technol       Date:  2005-06

2.  Low Oxygen Delivery as a Predictor of Acute Kidney Injury during Cardiopulmonary Bypass.

Authors:  Richard F Newland; Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2017-12

3.  Variation in Measurement and Reporting of Goal Directed Perfusion Parameters.

Authors:  Robert A Baker
Journal:  J Extra Corpor Technol       Date:  2017-06

4.  Evaluation of an autoregulatory ECMO system for total respiratory support in an acute ovine model.

Authors:  Robert G Conway; Zachary B Berk; Jiafeng Zhang; Tieluo Li; Douglas Tran; Zhongjun J Wu; Bartley P Griffith
Journal:  Artif Organs       Date:  2020-01-19       Impact factor: 3.094

5.  Development of a CO2 Sensor for Extracorporeal Life Support Applications.

Authors:  Michele Bellancini; Laura Cercenelli; Stefano Severi; Guido Comai; Emanuela Marcelli
Journal:  Sensors (Basel)       Date:  2020-06-27       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.