Limitations using the vacuum-assist venous drainage technique during cardiopulmonary bypass procedures.

David Jegger1, Hendrik T Tevaearai, Xavier M Mueller, Judith Horisberger, Ludwig K von Segesser.   

Abstract

Vacuum-assist venous drainage (VAVD) can increase venous blood return during cardiopulmonary bypass (CPB) procedures. However, the negative pressure created in the closed cardiotomy reservoir can be transmitted to the oxygenator if a nonocclusive or centrifugal arterial pump is used, resulting in bubble transgression (BT) from the gas to blood compartment of the oxygenator. We analyzed the vacuum pressure required to produce BT using an in vitro circuit including successively a closed reservoir, a pump (centrifugal or roller), and an oxygenator. A constant hydrostatic pressure was maintained onto the oxygenator. Vacuum was applied on the cardiotomy reservoir, progressively increasing negative pressure from 0 to -80 mmHg and monitoring BT with a bubble detector. Six different oxygenators were compared. A partially occlusive roller pump and a centrifugal pump were compared to a control, which was without any pump. A mean negative pressure of -53 +/- 7 mmHg was necessary to produce BT in all the oxygenators in the absence of a pump. The presence of a centrifugal pump between the reservoir and the oxygenator significantly increased the negative pressure required to produce BT compared to the control (-67 +/- 7 mmHg, p < .05). No bubbles were detected using the roller pump (> -80 mmHg needed for BT), thus statistically significant when compared to the centrifugal pump (p < .05). The centrifugal pump offers significant resistance to BT but not as much compared to the roller pump, though BT cannot be prevented if the pump is turned off while the vacuum remains on the reservoir. Therefore, VAVD is a safe technique as long as the perfusionist stops the vacuum when the arterial pump is no longer in use.

Entities:  

Mesh:

Year:  2003        PMID: 14653422

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  3 in total

Review 1.  Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass.

Authors:  Shigang Wang; Akif Undar
Journal:  J Extra Corpor Technol       Date:  2008-12

2.  [Improved quality of stored packed red blood cells by mechanical rinsing].

Authors:  F Münch; A Purbojo; F Wenzel; M Kohl; S Dittrich; M Rauh; R Zimmermann; N Kwapil
Journal:  Anaesthesiologie       Date:  2022-08-15

3.  Hydrodynamic Evaluations of Four Mock Femoral Venous Cannulas.

Authors:  Türker Şahin; Murat Tezer; Levent Cerit
Journal:  Braz J Cardiovasc Surg       Date:  2018 Sep-Oct
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.