Literature DB >> 14652837

A UK multi-centre experience with a second-generation endovascular stent-graft: results from the Zenith Users Group.

R J Hinchliffe1, J Goldberg, S T R Macsweeney.   

Abstract

BACKGROUND: Endovascular aneurysm repair has been criticised for high rates of technical failure and secondary intervention. Second generation stent-grafts have been developed in an attempt to reduce these problems. The results of a UK multi-centre experience with a second generation device (Zenith) are presented.
METHODS: Data were collected retrospectively from five experienced UK vascular centres in patients undergoing endovascular aneurysm repair with the Zenith stent-graft.
RESULTS: A total of 269 patients underwent attempted aneurysm repair with the Zenith device. Median aneurysm diameter was 65 (interquartile range 52-78) mm. There were no conversions to open repair. Peri-operative mortality was 4.1% (11/269). On the initial post-operative scan, 94.1% of aneurysms were successfully excluded. During a median follow-up of 363 (interquartile range 154-720) days there were 19 secondary interventions and two aneurysm ruptures.
CONCLUSIONS: Second generation endovascular stent-graft designs such as the Zenith are associated with low rates of intra-operative technical complications. Few secondary interventions have been necessary during follow-up; however, surveillance is essential to ensure they continue to perform.

Entities:  

Mesh:

Year:  2004        PMID: 14652837     DOI: 10.1016/j.ejvs.2003.10.004

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  2 in total

1.  Long-term outcomes of endovascular aortic aneurysm repair with the Zenith AAA endovascular graft: a single-center study.

Authors:  Hiroshi Kawamata; Hiroyuki Tajima; Tatsuo Ueda; Hidemasa Saito; Daisuke Yasui; Tadashi Kaneshiro; Naoko Takenoshita; Shouhei Mizushima; Takahiko Mine; Jiro Kurita; Yosuke Ishii; Tetsuro Morota; Takashi Nitta; Yuji Maruyama; Hajime Imura; Dai Nishina; Masahiro Fujii; Ryuzo Bessho
Journal:  Jpn J Radiol       Date:  2019-10-29       Impact factor: 2.374

2.  Fast and Accurate Computation of the Displacement Force of Stent Grafts after Endovascular Aneurysm Repair.

Authors:  Ming Qing; Zhan Liu; Tinghui Zheng
Journal:  Bioengineering (Basel)       Date:  2022-09-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.