Performance evaluation of a new compact hematology analyzer, the Sysmex pocH-100i.

C Briggs1, S Kunka, C Pennaneach, L Forbes, S J Machin.   

Abstract

Medical practitioners are searching for ways to improve the quality and outcome of care while decreasing cost. One technological advance that may facilitate this is point-of-care testing. The pocH-100i hematology analyzer (Sysmex Corporation, Kobe, Japan) is a compact device designed specifically for point-of-care testing environments. The analyzer provides a full blood count and a 3-part differential leukocyte count. A distinct advantage is that neutrophils are reported separately, which is useful for monitoring oncology patients. The device was evaluated in accordance with International Council for Standardization in Haematology (ICSH) guidelines for precision, linearity, carryover, and effects of sample aging. The pocH-100i analyzer was compared with the Sysmex KX-21N device. The results for all parameters tested were almost identical. When samples including blast cells, immature granulocytes, and nucleated red blood cells were excluded, the pocH-100i automated differential compared well with a 400-cell manual differential. Results for neutrophils (r2 = 0.996), lymphocytes (r2 = 0.999), and the "mixed" population of cells (r2 = 0.611) indicated the pocH-100i analyzer would be highly suitable for low-volume laboratories and near-patient services.

Entities:  

Mesh:

Year:  2003        PMID: 14649465

Source DB:  PubMed          Journal:  Lab Hematol        ISSN: 1080-2924


  3 in total

Review 1.  Point-of-Care-Testing in Acute Stroke Management: An Unmet Need Ripe for Technological Harvest.

Authors:  Dorin Harpaz; Evgeni Eltzov; Raymond C S Seet; Robert S Marks; Alfred I Y Tok
Journal:  Biosensors (Basel)       Date:  2017-08-03

2.  Impact of Interfacing Near Point of Care Clinical Chemistry and Hematology Analyzers at Urgent Care Clinics at an Academic Health System.

Authors:  Zachary J Fleishhacker; Prerna Rastogi; Scott R Davis; Dean R Aman; Cory S Morris; Richard L Dyson; Matthew D Krasowski
Journal:  J Pathol Inform       Date:  2022-01-20

3.  Measurement of CD4+ T cells in point-of-care settings with the Sysmex pocH-100i haematological analyser.

Authors:  C Briggs; S Machin; M Müller; W Haase; K Hofmann; F Forstreuter; R Hinzmann
Journal:  Int J Lab Hematol       Date:  2007-12-20       Impact factor: 2.877

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.