Literature DB >> 14648537

Evaluation of a vessel-sealing device for small intestinal resection and anastomosis in normal horses.

Marilyn L Rumbaugh1, Daniel J Burba, Claudio Natalini, Giselle Hosgood, Rustin M Moore.   

Abstract

OBJECTIVES: To compare arterial bursting pressure after vessel closure using a vessel-sealing device (LigaSure Atlas Laparoscopic Sealer/Divider Instrument; Valleylab, Boulder, CO), a ligate-and-divide stapling device (LDS), and 2-0 polydioxanone suture. To evaluate the LigaSure Atlas as a method for ligation of the mesenteric vasculature during small intestinal resection in normal horses. STUDY
DESIGN: Experimental study. ANIMALS: Part A: jejunal segments from 19 horses. Part B: 6 horses, aged 1 to 18 years, weighing 330 to 509 kg.
METHODS: Part A: Jejunal segments with mesenteric vessels were collected from 19 horses. After closure by 1 of 3 methods (LigaSure Atlas, LDS, 2-0 polydioxanone) arteries were cannulated, and bursting pressure was measured by incrementally increasing intraluminal pressure until failure. Part B: Six horses had jejunal resection and anastomosis using a vessel-sealing device (LigaSure Atlas) to provide hemostasis of the mesenteric vasculature. Horses were monitored clinically for 4 weeks.
RESULTS: Part A: Mean +/- SEM bursting pressure after 2-0 polydioxanone ligation (1,014.50 +/- 279.05 mm Hg) was significantly greater than mean bursting pressure after LigaSure (554.25 +/- 228.79 mm Hg), which was significantly greater than the mean bursting pressure after LDS (373.25 +/- 183.69). Part B: No major operative or postoperative hemorrhage occurred after application of the LigaSure Atlas for sealing and transecting mesenteric vasculature during small intestinal resection in normal horses.
CONCLUSIONS: The LigaSure Atlas appears to be a safe method for hemostasis of the mesenteric vasculature during small intestinal resection in normal horses. CLINICAL RELEVANCE: Benefits of the LigaSure Atlas vessel-sealing device include reduced time required to provide hemostasis, acceptable arterial bursting pressure, no remaining foreign material, and no risk for ligature slippage. In our experience, use of the LigaSure Atlas during small intestinal resection and anastomosis in horses is safe for ligation of vessels less than or equal to 7-mm diameter. Copyright 2003 by The American College of Veterinary Surgeons

Entities:  

Mesh:

Year:  2003        PMID: 14648537     DOI: 10.1111/j.1532-950x.2003.00574.x

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  4 in total

1.  Evaluation of the LigaSure(™) Vessel Sealing System for bowel transection and intestinal anastomosis-an experimental study in a porcine model.

Authors:  Tobias Gehrig; A T Billeter; A L Wekerle; M Shevchenko; K Brand; B P Müller-Stich
Journal:  Langenbecks Arch Surg       Date:  2016-03-23       Impact factor: 3.445

2.  Parallel alternating sliding knots are effective for ligation of mesenteric arteries during resection and anastomosis of the equine jejunum.

Authors:  Marco Gandini; Gessica Giusto; Francesco Comino; Eleonora Pagliara
Journal:  BMC Vet Res       Date:  2014-07-07       Impact factor: 2.741

3.  Ex-Vivo Evaluation of "First Tip Closing" Radiofrequency Vessel Sealing Devices for Swine Small Intestinal Transection.

Authors:  Luca Lacitignola; Alberto Crovace; Giuseppe Passantino; Francesco Staffieri
Journal:  Vet Sci       Date:  2022-08-19

4.  Ex vivo comparison of sliding knot ligatures vs. haemostatic clips for equine small intestinal mesenteric vessel occlusion.

Authors:  Gessica Giusto; Marco Gandini
Journal:  BMC Vet Res       Date:  2020-08-12       Impact factor: 2.741

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.