OBJECTIVE: To compare the fixation rigidity of recently developed external fixation systems (EFSs) to that of the traditional Kirschner-Ehmer (KE) system. STUDY DESIGN: In vitro biomechanical study. SAMPLE POPULATION: Five different EFSs (KE, Secur-U, small SK carbon fiber, small SK titanium, large SK carbon fiber) were assembled into 7 frame geometries to stabilize Delrin plastic rods with a 1-cm gap. METHODS: External skeletal fixation (ESF) constructs were tested in axial compression, torsion, medial-lateral bending, and cranial-caudal bending. Testing was conducted within the elastic range of each fixator. Mean stiffness in each mode was determined from the slope of the linear portion of the load-deformation curve. Comparison of stiffness values of each EFS within each loading mode and frame type was performed with 1-way analysis of variance (P <.05). RESULTS: Mean stiffness values were significantly higher for the large SK EFS in all frame types compared with KE but were equal in torsional stiffness in the double-bar type 1a frame. The small SK EFS with titanium connecting bar had greater stiffness than the KE in all modes for frame types Ia, Ia-accessory bar, and II-modified. No overall difference was detected between the KE EFS and the small SK with carbon fiber rod. The stiffness of the Secur-U type Ia frame with augmentation plate was significantly greater than the KE type Ia with accessory bar. CONCLUSIONS: The newer external fixation systems evaluated in this study provided fixation rigidity equal to or greater than that of the KE system. CLINICAL RELEVANCE: EFSs with increased frame rigidity should permit the use of less complex frame designs while providing fracture stability. Copyright 2003 by The American College of Veterinary Surgeons
OBJECTIVE: To compare the fixation rigidity of recently developed external fixation systems (EFSs) to that of the traditional Kirschner-Ehmer (KE) system. STUDY DESIGN: In vitro biomechanical study. SAMPLE POPULATION: Five different EFSs (KE, Secur-U, small SK carbon fiber, small SK titanium, large SK carbon fiber) were assembled into 7 frame geometries to stabilize Delrin plastic rods with a 1-cm gap. METHODS: External skeletal fixation (ESF) constructs were tested in axial compression, torsion, medial-lateral bending, and cranial-caudal bending. Testing was conducted within the elastic range of each fixator. Mean stiffness in each mode was determined from the slope of the linear portion of the load-deformation curve. Comparison of stiffness values of each EFS within each loading mode and frame type was performed with 1-way analysis of variance (P <.05). RESULTS: Mean stiffness values were significantly higher for the large SK EFS in all frame types compared with KE but were equal in torsional stiffness in the double-bar type 1a frame. The small SK EFS with titanium connecting bar had greater stiffness than the KE in all modes for frame types Ia, Ia-accessory bar, and II-modified. No overall difference was detected between the KE EFS and the small SK with carbon fiber rod. The stiffness of the Secur-U type Ia frame with augmentation plate was significantly greater than the KE type Ia with accessory bar. CONCLUSIONS: The newer external fixation systems evaluated in this study provided fixation rigidity equal to or greater than that of the KE system. CLINICAL RELEVANCE: EFSs with increased frame rigidity should permit the use of less complex frame designs while providing fracture stability. Copyright 2003 by The American College of Veterinary Surgeons