Literature DB >> 14645877

Biologic response to polymer-coated stents: in vitro analysis and results in an iliac artery sheep model.

Karl Schürmann1, Joerg Lahann, Pascal Niggemann, Bernd Klosterhalfen, Julius Meyer, Arthur Kulisch, Doris Klee, Rolf W Günther, Dierk Vorwerk.   

Abstract

PURPOSE: To evaluate biologic response to poly(hydroxymethyl-p-xylylene-co-p-xylylene) (PHPX)-coated stents in vitro and in vivo in sheep.
MATERIALS AND METHODS: Physical stability, hemocompatibility, and cytotoxicity of the coating were first assessed in vitro. Thirty-six self-expanding nitinol (Memotherm), 24 stainless steel balloon-mounted (Palmaz), and 12 self-expanding nitinol (ZA) stents were coated with PHPX by using chemical vapor deposition polymerization. Seventy-two coated and 72 uncoated stents were placed into iliac arteries of 36 sheep. Sheep were classified into three groups of 12 animals each. In each group, six sheep were killed after 1 month; six, after 6 months. In each sheep, two uncoated stents were placed into one limb; two coated stents of the same type, into the opposite limb. In groups 1 and 2, Palmaz and Memotherm stents were used; in group 3, Memotherm and ZA stents were used. In groups 1 and 3, arteries were healthy. In group 2, arteries were pretreated with a Fogarty maneuver. Stent patency was measured with intravascular ultrasonography (US) and histologic analysis. Cellular response to coated and uncoated stents was assessed. Measurements were compared (Wilcoxon test).
RESULTS: In vitro, PHPX coating was stable; hemocompatibility and cytotoxicity were similar to those of stainless steel. In vivo, patency of coated and uncoated Palmaz and ZA stents was not different (P >.05). Patency of coated and uncoated Memotherm stents did not differ in four of six follow-up subgroups, but it was significantly reduced in group 2 after 6 months (intravascular US, P =.03; histologic analysis, P =.01) and in group 3 after 1 month (histologic analysis, P =.01). Histologically, the cellular response to coated and uncoated stents was not different (P >.05).
CONCLUSION: PHPX coating had good physical stability and biocompatibility in vitro and in vivo. Performance of coated and uncoated Palmaz and ZA stents was similar. Patency of Memotherm stents was similar in four of six follow-up subgroups. Materials effects did not result in severely enhanced neointimal hyperplasia. Copyright RSNA, 2003

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14645877     DOI: 10.1148/radiol.2301021071

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Cellular transduction gradients via vapor-deposited polymer coatings.

Authors:  Yaseen M Elkasabi; Joerg Lahann; Paul H Krebsbach
Journal:  Biomaterials       Date:  2010-12-22       Impact factor: 12.479

2.  Granulation encapsulated stent: a new therapeutic approach for vascular implantation.

Authors:  Lilong Tang; Xiaochao Chen; Shuangbo Tang; Thomas LaLonde; Julius M Gardin
Journal:  Heart       Date:  2006-08-16       Impact factor: 5.994

3.  Vapor sublimation and deposition to build porous particles and composites.

Authors:  Hsing-Ying Tung; Zhen-Yu Guan; Ting-Yu Liu; Hsien-Yeh Chen
Journal:  Nat Commun       Date:  2018-07-02       Impact factor: 14.919

  3 in total