Structural changes in silicon rubber peritoneal dialysis catheters in patients using mupirocin at the exit site.

M Khandelwal1, S Bailey, S Izatt, M Chu, S Vas, J Bargman, D Oreopoulos.   

Abstract

Structural damage to polyurethane PD (peritoneal dialysis) catheters in patients using mupirocin ointment is widely appreciated, but damage to silicon rubber PD catheters is less well described. Ten catheters (6.6%) out of 152 were found to have structural alterations such as opacification, ballooning, thinning, and rupture. The duration of PD in these 10 patients ranged from 23 months to 80 months (mean duration 51.1 months). The frequency of mupirocin application varied from daily (2 cases) to 2-3 times per week (7 cases). In eight catheters opacification occurred at the exit site whereas one catheter showed opacification midway between the exit site and the titanium adaptor. One catheter showed opacification, ballooning, and thinning at the exit site ruptured in the form of two slit-like openings. In conclusion, various structural changes such as opacification, ballooning or thinning were seen in 6.6% of silicon rubber PD catheters in patients using mupirocin at the exit site. Although the mechanism remains elusive, mupirocin or the antiseptic solution alone or in combination may be contributory. We believe that this is an under-reported complication and encourage other health care givers to incorporate a search for such changes during clinic visits.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14636007     DOI: 10.1177/039139880302601007

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  3 in total

1.  Exit-site infection and acute peritonitis due to peritoneal dialysis catheter rupture.

Authors:  Maša Knehtl; Robert Ekart; Sebastjan Bevc; Matjaž Horvat; Radovan Hojs
Journal:  Perit Dial Int       Date:  2015 May-Jun       Impact factor: 1.756

2.  Drainage failure because of spontaneous fracture of the peritoneal dialysis catheter.

Authors:  H R Kim; M J Lee; J E Song; J H Han; T H Yoo; S W Kang; K H Choi; S H Han
Journal:  Perit Dial Int       Date:  2013 Mar-Apr       Impact factor: 1.756

3.  A material stress test study on occurrence of leakage and material failure of peritoneal dialysis (PD) catheters.

Authors:  Matthias Klingele; Martin Carstens; Lea Baerens; Matthias W Laschke; Wolfgang Metzger; Danilo Fliser; Clemens M Meier
Journal:  Sci Rep       Date:  2021-05-13       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.