Paraspinal electromyography that compares concentric with monopolar needles: a blinded study.

Henry C Tong1, Irene A Young, John Koch, Andrew J Haig, Karen S J Yamakawa, Agnes Wallbom.   

Abstract

OBJECTIVE: To establish interrater reliability for paraspinal muscle needle electromyography study with both monopolar and concentric needles in symptomatic and asymptomatic persons and to further establish normative data for paraspinal needle study.
DESIGN: At a university spine center, participants with and without radiating low back pain were evaluated with the mini-paraspinal mapping paraspinal needle technique by an unblinded and a blinded electromyographer.
RESULTS: In the symptomatic group, the intraclass correlation coefficient between concentric and monopolar needles was 0.793; between monopolar needles, it was 0.876; and between concentric needles, it was 0.966. In the asymptomatic group, the mean total score was 0.25.
CONCLUSIONS: The good interrater reliability with the same needle type helps support the validity of the needle electromyography study of the paraspinal muscles. The good correlation between the concentric and monopolar needles shows the data published using monopolar needle data also apply to studies using paraspinal needle electromyography with concentric needles. The low score with the asymptomatic group reaffirms that using a cutoff score of >2 as abnormal has a false-positive rate of <5%.

Entities:  

Mesh:

Year:  2003        PMID: 14627928     DOI: 10.1097/01.PHM.0000091983.14237.00

Source DB:  PubMed          Journal:  Am J Phys Med Rehabil        ISSN: 0894-9115            Impact factor:   2.159


  2 in total

1.  Diagnostic accuracy of multifidus muscle spontaneous activity by needle electromyography for the detection of lumbar foraminal and lateral exit-zone stenosis.

Authors:  Mikinobu Takeuchi; Norimitsu Wakao; Mitsuhiro Kamiya; Masahiko Gosho; Koji Osuka; Atsuhiko Hirasawa; Aichi Niwa; Masahiro Aoyama; Reo Kawaguchi; Hiroshi Shima; Masakazu Takayasu
Journal:  Eur Spine J       Date:  2015-03-04       Impact factor: 3.134

2.  A prospective, masked 18-month minimum follow-up on neurophysiologic changes in persons with spinal stenosis, low back pain, and no symptoms.

Authors:  Andrew J Haig; Karen S J Yamakawa; Christopher Parres; Anthony Chiodo; Henry Tong
Journal:  PM R       Date:  2009-02-03       Impact factor: 2.298

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.