Syringe calibration factors for the NPL Secondary Standard Radionuclide Calibrator for selected medical radionuclides.

D K Tyler1, M J Woods.   

Abstract

Before a radiopharmaceutical is administered to a patient, its activity needs to be accurately assayed. This is normally done via a radionuclide calibrator, using a glass vial as the calibration device. The radionuclide is then transferred to a syringe and it is now becoming common practice to re-measure the syringe and use this value as the activity administered to the patient. Due to elemental composition and geometrical differences, etc. between the glass vial and the syringe, the calibration factors are different for the two containers and this can lead to an incorrect activity being given to the patient unless a correction is applied for these differences. To reduce the uncertainty on syringe measurements, syringe calibration factors and volume correction factors for the NPL Secondary Standard Radionuclide Calibrator have been derived by NPL for several medically important radionuclides. It was found that the differences between the calibration factors for the syringes and glass vials depend on the energies of the photon emissions from the decay of the radionuclides; the lower the energy, the greater the difference. As expected, large differences were observed for 125I (70%) and only small differences for 131I. However, for radionuclides such as 99mTc and 67Ga, differences of up to 30% have been observed. This work has shown the need for the use of specifically derived syringe calibration factors as well as highlighting the complexity of the problem with regard to syringe types, procurement, etc.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14622937     DOI: 10.1016/j.apradiso.2003.08.002

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  2 in total

1.  Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters.

Authors:  Sarika Sharma; Baljinder Singh; Ashwani Koul; Bhagwant Rai Mittal
Journal:  Indian J Nucl Med       Date:  2015 Apr-Jun

2.  Assessment of the relative contribution of volume and concentration changes in Yttrium-90 labelled resin microspheres on ionization chamber measurements.

Authors:  Nicholas Forwood; Kathy P Willowson; Michael Tapner; Dale L Bailey
Journal:  Australas Phys Eng Sci Med       Date:  2017-11-17       Impact factor: 1.430

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.