Literature DB >> 14620981

Poor experience with a hinged endoprosthesis (WEKO) for the metacarpophalangeal joints: all 28 prostheses removed within 2 years in 8 patients having rheumatoid arthritis.

Sebastian Radmer1, Reimer Andresen, Martin Sparmann.   

Abstract

We prospectively assessed the outcome of implanting a hinged prosthesis in destroyed metacarpophalangeal (MCP) joints (Larsen stage III-V). We implanted 28 cementless, axis-coupled total endoprostheses with hinging (WEKO prosthesis, Implant-Service GmbH, Hamburg, Germany) in 8 women (mean age 62 (47-80) years) suffering from rheumatoid arthritis. The mean follow-up period was 15 (12-18) months, and the evaluation was based on objective criteria, including joint motion, ulnar deviation of the long fingers, grip strength, radiographic migration and torsion of the prostheses, as well as the patients' subjective satisfaction. 12 months postoperatively, the mean arc of flexion was 30 (22-35) degrees, and the mean extension lag was 43 (40-48) degrees. Although no ulnar deviation was seen in 2 fingers, it was < 10 degrees in 3, between 10 degrees and 20 degrees in 7 of the fingers, and more than 20 degrees in 16. None of the patients could clench their fist firmly. We found prosthesis migration in 20 and torsion in 19 fingers. 2 years postoperatively, we had to remove all of the prostheses due to functional failure.

Entities:  

Mesh:

Year:  2003        PMID: 14620981     DOI: 10.1080/00016470310018009

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  1 in total

1.  Wear testing of a DJOA finger prosthesis in vitro.

Authors:  Thomas J Joyce
Journal:  J Mater Sci Mater Med       Date:  2010-02-11       Impact factor: 3.896

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.