Literature DB >> 14604723

Effect of infusion bottle height on lens power after lens refilling with and without a plug.

Steven A Koopmans1, Thom Terwee, Henk J Haitjema, Jan Barkhof, Aart C Kooijman.   

Abstract

PURPOSE: To evaluate the influence of intraoperative infusion bottle height on the power of refilled pig lenses.
SETTING: Research Laboratory, Pharmacia Intraocular Lens Manufacturing Plant, Groningen, The Netherlands.
METHODS: This study comprised 2 groups of pig eyes. In 1 group, the lens was refilled with silicone oil using a plug to close the capsulorhexis; in the other group, no plug was used. The anterior chamber depth, lens thickness, and vitreous chamber depth were measured by A-scan ultrasound. The total refraction was measured with a Hartinger refractometer. Measurements were performed with the infusion bottle at 0 cm, 25 cm, 50 cm, 75 cm, and 100 cm above eye level. Calculations using a model eye were performed to evaluate the change in lens power based on empirical data.
RESULTS: The mean change in the power of refilled lenses caused by varying the height of the infusion bottle was 1.8 diopters. Lenses refilled with a plug had a thickness similar to that of natural lenses. Lenses refilled without a plug were significantly thinner (P<.05). The power of lenses refilled with a plug was significantly higher that of lenses refilled without a plug (P<.05).
CONCLUSIONS: During lens refilling, infusion bottle height influenced the resulting lens power. Using a plug to close the capsulorhexis resulted in refilled lens dimensions similar to those of the natural lens.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14604723     DOI: 10.1016/s0886-3350(03)00356-0

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

Review 1.  Accommodating intraocular lenses: a critical review of present and future concepts.

Authors:  R Menapace; O Findl; K Kriechbaum; Ch Leydolt-Koeppl
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-08-30       Impact factor: 3.117

2.  [Lens refilling].

Authors:  H-J Hettlich
Journal:  Ophthalmologe       Date:  2010-05       Impact factor: 1.059

  2 in total