| Literature DB >> 14594059 |
Jorge E Carranza1, Emily Gibb, Ben W Smith, David W Hahn, James D Winefordner.
Abstract
The performance and sensitivity of an intensified CCD array system and a nonintensified CCD array detector system are compared for laser-induced breakdown spectroscopy (LIBS). LIBS measurements were recorded in a calcium-based aerosol-seeded gas stream at ambient pressure. The signal-to-noise ratio based on the 393.37-nm calcium emission line was calculated as a function of detector delay with respect to the plasma-initiating laser pulse. Both ensemble-averaging and single-shot spectral analyses were performed. For all conditions, the intensified CCD system provided an enhanced signal-to-noise ratio compared with the nonintensified CCD system.Entities:
Year: 2003 PMID: 14594059 DOI: 10.1364/ao.42.006016
Source DB: PubMed Journal: Appl Opt ISSN: 1559-128X Impact factor: 1.980