The effect of the Varian amorphous silicon electronic portal imaging device on exit skin dose.

Warren Kilby1, Clare Savage.   

Abstract

Measurements have been made of the increase in exit surface dose resulting from backscattered radiation generated by the Varian amorphous silicon electronic portal imaging device (EPID). An increase of < or = 14% was demonstrated at both 6 MV and 10 MV, in a manner which suggests that backscatter from the EPID acts to re-establish electronic equilibrium at the exit surface, normally absent in the build-down region. The magnitude of this effect was influenced by field size, measurement depth and exit surface to EPID distance. Assuming typical constraints of portal imaging frequency and geometry, the results suggest that EPID generated backscatter is unlikely to alter the frequency or severity of exit skin reactions. However, the results do suggest that a limit on the minimum separation between the EPID and the exit surface should be set, and that similar investigations should be made for other EPID models.

Mesh:

Substances:

Year:  2003        PMID: 14579855     DOI: 10.1088/0031-9155/48/19/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  An initial study on the estimation of time-varying volumetric treatment images and 3D tumor localization from single MV cine EPID images.

Authors:  Pankaj Mishra; Ruijiang Li; Raymond H Mak; Joerg Rottmann; Jonathan H Bryant; Christopher L Williams; Ross I Berbeco; John H Lewis
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

2.  Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors.

Authors:  J Rottmann; P Keall; R Berbeco
Journal:  Phys Med Biol       Date:  2013-05-28       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.