Literature DB >> 1457954

An electrically powered total artificial heart. Over 1 year survival in the calf.

A J Snyder1, G Rosenberg, J Reibson, J H Donachy, G A Prophet, J Arenas, B Daily, S McGary, O Kawaguchi, R Quinn.   

Abstract

An electric motor driven orthotopic artificial heart was implanted in a 110 kg female Holstein calf as part of a series of 12 such implants intended to demonstrate the in vivo durability and compatibility of the device. The device uses pusher plates set into motion by a reversing brushless DC motor and roller screw to alternately eject two cylindrical sac type blood pumps. The pumps use Bjork-Shiley Delrindisc convexo-concave or monostrut valves. The left pump provides an 88-90 ml dynamic stroke volume. Woven Dacron grafts and polyurethane coated Dacron/Lycra cuffs are used to attach the device to the major arteries and atria, respectively. A polyurethane conduit and anchoring skin button bring motor wires percutaneously to an extracorporeal controller. The controller provides balanced cardiac output sensitive to atrial or aortic pressures, without operator intervention. The system is hermetically sealed and uses a simple compliance sac to maintain thoracic pressure between the pumps. The calf recovered uneventfully from surgery and thrived thereafter. She was killed on the 388th post-operative day because of worsening cardiac insufficiency. The previous three operative survivors in this series lived 131, 134, and 204 days. These results indicate the device's good potential for durability and body compatibility.

Entities:  

Mesh:

Year:  1992        PMID: 1457954     DOI: 10.1097/00002480-199207000-00130

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  5 in total

Review 1.  Development of mechanical circulatory support devices at the University of Tokyo.

Authors:  Yusuke Abe; Takashi Isoyama; Itsuro Saito; Shuichi Mochizuki; Minoru Ono; Hidemoto Nakagawa; Noriyuki Taniguchi; Norihiko Mitsumune; Ayaka Sugino; Mie Mitsui; Koki Takiura; Toshiya Ono; Akimasa Kouno; Tsuneo Chinzei; Shinichi Takamoto; Kou Imachi
Journal:  J Artif Organs       Date:  2007-06-20       Impact factor: 1.731

Review 2.  Total artificial hearts: past, present, and future.

Authors:  William E Cohn; Daniel L Timms; O H Frazier
Journal:  Nat Rev Cardiol       Date:  2015-06-02       Impact factor: 32.419

3.  Concept of left atrial pressure estimation using its pulsatile amplitude in the helical flow total artificial heart.

Authors:  Sheng-Yuan Wu; Itsuro Saito; Takashi Isoyama; Yusuke Inoue; Masami Sato; Shintaro Hara; Xin-Yang Li; Terumi Yurimoto; Haruka Murakami; Yukino Kawase; Toshiya Ono; Yusuke Abe
Journal:  J Artif Organs       Date:  2014-09-03       Impact factor: 1.731

Review 4.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

5.  A nonpulsatile total artificial heart with 1/R control.

Authors:  Yusuke Abe; Itsuro Saito; Takashi Isoyama; Hidekazu Miura; Wei Shi; Sachiko Yamaguchi; Yusuke Inoue; Hidemoto Nakagawa; Minoru Ono; Ayumi Kishi; Toshiya Ono; Akimasa Kouno; Tsuneo Chinzei; Kou Imachi
Journal:  J Artif Organs       Date:  2008-12-17       Impact factor: 1.731

  5 in total