Literature DB >> 1457938

Thrombin activity resides on LVAD Dacron inflow and outflow grafts.

W R Wagner1, P C Johnson, B V Heil, K A Thompson, R L Kormos, B P Griffith.   

Abstract

Patients on left ventricular assist devices (LVADs) are at increased risk for thromboembolism, and they experience elevations in platelet release and thrombin activity indices during device implantation. The Dacron grafts that lead from the ventricle to the LVAD and back to the aorta may harbor thrombin activity and protect this thrombin from anticoagulant action. To investigate this possibility, specimens isolated from LVAD grafts at the time of device explantation or implantation were incubated with citrated platelet poor plasma (cPPP), cPPP and 2 U/ml of heparin, cPPP and 2 U/ml of recombinant hirudin (r-hirudin), or cPPP and 50 U/ml of heparin. Thrombin activity was measured by the increase in incubated cPPP fibrinopeptide A (FPA) concentrations over cPPP FPA levels, without material contact. A chromogenic substrate for thrombin verified the effects seen. Thrombin activity was found at comparable levels on both the inflow and outflow LVAD grafts at explantation. This activity was not inhibited by low concentrations of heparin, but 50 U/ml of heparin and 2 U/ml of r-hirudin reduced the activity. At graft implantation (after preclotting), thrombin activity was higher than at explantation, but susceptibility to inhibition by heparin was also significantly greater. These results confirm that the LVAD Dacron grafts harbor surface thrombin activity resistant to anticoagulation that may be a primary source for LVAD thrombogenicity.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1457938     DOI: 10.1097/00002480-199207000-00114

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  3 in total

1.  Elevated Angiopoietin-2 Level in Patients With Continuous-Flow Left Ventricular Assist Devices Leads to Altered Angiogenesis and Is Associated With Higher Nonsurgical Bleeding.

Authors:  Corey E Tabit; Phetcharat Chen; Gene H Kim; Savitri E Fedson; Gabriel Sayer; Mitchell J Coplan; Valluvan Jeevanandam; Nir Uriel; James K Liao
Journal:  Circulation       Date:  2016-06-28       Impact factor: 29.690

Review 2.  Miniaturization of mechanical circulatory support systems.

Authors:  Guruprasad A Giridharan; Thomas J Lee; Mickey Ising; Michael A Sobieski; Steven C Koenig; Laman A Gray; Mark S Slaughter
Journal:  Artif Organs       Date:  2012-08       Impact factor: 3.094

Review 3.  Perioperative pharmacotherapy in patients with left ventricular assist devices.

Authors:  Nicholas C Dang; Yoshifumi Naka
Journal:  Drugs Aging       Date:  2004       Impact factor: 4.271

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.