Literature DB >> 14578312

Critique of the Guide to the expression of uncertainty in measurement method of estimating and reporting uncertainty in diagnostic assays.

Jan S Krouwer1.   

Abstract

BACKGROUND: The Guide to the Expression of Uncertainty in Measurement (GUM) provides instructions for constructing uncertainty intervals for a measurement. This method is usually reserved for reference materials, but GUM has been recently proposed as a way to express uncertainty for commercial diagnostic assays.
METHODS: Using the official GUM standard and published applications of GUM to commercial diagnostic assays, I undertook an analysis to evaluate whether applying GUM to commercial diagnostic assays is warranted.
RESULTS: Certain important assays, such as troponin I, would not be candidates for GUM because troponin I is not a well-defined physical quantity. Unlike definitive methods, in which efforts are taken to detect and eliminate all systematic error sources, commercial assays often trade off features such as ease of use and cost with accuracy and allow systematic errors to be present as long as the overall accuracy meets the medical need goal. Laboratories are hindered in preparing GUM models because the knowledge required to specify some systematic errors is often available only to manufacturers. Some non-GUM methods to estimate uncertainty rely on observed data, which include both known and unknown sources of error. The occurrence of large, unknown errors for assays in routine use (e.g., outliers) is not unusual because diagnostic assays must be chemically specific in the presence of thousands of potentially interfering substances. There is no provision in GUM to deal with unexplained outliers, which may lead to uncertainty intervals that are not wide enough. Some clinicians assume that diagnostic assay results have little uncertainty. This situation may be made worse by including an uncertainty interval, which implies certification.
CONCLUSIONS: Evaluations for accuracy (total analytical error) based on describing the distribution of result differences between commercial assays and reference methods indicate that some assays have a few results with large differences (e.g., outliers). This leads to a wide accuracy interval (total analytical error limits). It is unlikely that GUM would be able to predict these wide intervals, especially because there is little or no provision for outlier treatment in GUM. Presenting too narrow GUM uncertainty intervals to clinicians would be misleading. The modeling used by practitioners of the GUM method is potentially useful in improving quality, but commercial diagnostic assays are not ready for GUM uncertainty statements.

Mesh:

Substances:

Year:  2003        PMID: 14578312     DOI: 10.1373/clinchem.2003.019505

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  6 in total

1.  Uncertainty of measurement in quantitative medical testing: a laboratory implementation guide.

Authors:  G H White; I Farrance
Journal:  Clin Biochem Rev       Date:  2004

Review 2.  Uncertainty in measurement: a review of monte carlo simulation using microsoft excel for the calculation of uncertainties through functional relationships, including uncertainties in empirically derived constants.

Authors:  Ian Farrance; Robert Frenkel
Journal:  Clin Biochem Rev       Date:  2014-02

3.  Disregarded Measurement Uncertainty Contributions and Their Magnitude in Measuring Plasma Glucose.

Authors:  Astrid Petersmann; Rainer Macdonald; Matthias Nauck
Journal:  J Diabetes Sci Technol       Date:  2020-11-20

4.  Two evaluation budgets for the measurement uncertainty of glucose in clinical chemistry.

Authors:  Hui Chen; Ling Zhang; Xiaoyun Bi; Xiaoling Deng
Journal:  Korean J Lab Med       Date:  2011-06-28

5.  Analytical robustness of nine common assays: frequency of outliers and extreme differences identified by a large number of duplicate measurements.

Authors:  Stefanie Neubig; Anne Grotevendt; Anders Kallner; Matthias Nauck; Astrid Petersmann
Journal:  Biochem Med (Zagreb)       Date:  2017-02-15       Impact factor: 2.313

6.  Spatial and temporal patterns of carbon storage in forest ecosystems on Hainan island, southern China.

Authors:  Hai Ren; Linjun Li; Qiang Liu; Xu Wang; Yide Li; Dafeng Hui; Shuguang Jian; Jun Wang; Huai Yang; Hongfang Lu; Guoyi Zhou; Xuli Tang; Qianmei Zhang; Dong Wang; Lianlian Yuan; Xubing Chen
Journal:  PLoS One       Date:  2014-09-17       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.