Literature DB >> 14577011

Clinical evaluation of stereotactic brain biopsies with an MKM-mounted instrument holder.

P W A Willems1, H J Noordmans, L M P Ramos, M J B Taphoorn, J W Berkelbach van der Sprenkel, M A Viergever, C A F Tulleken.   

Abstract

OBJECT: The aim of this study was to assess the clinical usefulness and accuracy of robot-assisted frameless stereotactic brain biopsies with a recently introduced MKM-mounted instrument holder.
METHODS: Twenty-three patients with intracranial lesions participated in this study. Depending on the size of the intracranial lesion, fiducials for image-to-patient co-ordinate transformation consisted either of bone screws or adhesive markers. Shortly after surgery, postoperative MRI-imaging was performed to demonstrate the location of the biopsy site. These images were compared with the preoperative images to assess the biopsy localisation error.
RESULTS: Postoperative biopsy sites could be demonstrated in six patients with bone screws and in 14 with adhesive markers. These two subgroups yielded average biopsy localisation errors of 3.3 mm (SD 1.7 mm) and 4.5 mm (SD 2.0 mm) respectively. This difference was not statistically significant. One biopsy was located in a liquefied haematoma. All others yielded pathological tissue. There were two postoperative haemorrhages, of which only one was temporarily symptomatic. There was no mortality in the first 30 days after surgery.
CONCLUSIONS: Robot-assisted frameless point-stereotactic techniques represent an alternative to frame-based techniques for the performance of stereotactic biopsies.

Entities:  

Mesh:

Year:  2003        PMID: 14577011     DOI: 10.1007/s00701-003-0112-4

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  5 in total

1.  Laser-assisted flat-detector CT-guided intracranial access.

Authors:  Daniel L Cooke; Michael R Levitt; Louis J Kim; Danial K Hallam; Laligam N Sekhar; Basavaraj V Ghodke
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-08-05       Impact factor: 2.924

Review 2.  Neuronavigation and surgery of intracerebral tumours.

Authors:  P W A Willems; J W Berkelbach van der Sprenkel; C A F Tulleken; M A Viergever; M J B Taphoorn
Journal:  J Neurol       Date:  2006-09-20       Impact factor: 4.849

3.  Robot-assisted stereotactic brainstem biopsy in children: prospective cohort study.

Authors:  William Dawes; Hani J Marcus; Martin Tisdall; Kristian Aquilina
Journal:  J Robot Surg       Date:  2018-12-06

4.  A Comparation Between Frame-Based and Robot-Assisted in Stereotactic Biopsy.

Authors:  Yue Hu; Pu Cai; Huawei Zhang; Aihemaitiniyazi Adilijiang; Jun Peng; Yun Li; Shanli Che; Fei Lan; Changqing Liu
Journal:  Front Neurol       Date:  2022-07-18       Impact factor: 4.086

5.  Robot-assisted stereotactic brain biopsy: systematic review and bibliometric analysis.

Authors:  Hani J Marcus; Vejay N Vakharia; Sebastien Ourselin; John Duncan; Martin Tisdall; Kristian Aquilina
Journal:  Childs Nerv Syst       Date:  2018-05-10       Impact factor: 1.475

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.