Measurement of the ultrasound backscatter signal from three seed types as a function of incidence angle: application to permanent prostate brachytherapy.

Brian J Davis1, Randall R Kinnick, Mostafa Fatemi, Eugene P Lief, Richard A Robb, James F Greenleaf.   

Abstract

PURPOSE: To measure the relative ultrasound backscatter of different seed types as a function of seed orientation and to evaluate the corresponding images of these seeds. METHODS AND MATERIALS: Three seed types were evaluated: OncoSeed (standard), EchoSeed (corrugated), and RAPID Strand(RS). Ultrasound images for angles of incidence varying from 90 degrees (perpendicular) to 20 degrees at 5MHz and 7.5MHz were produced by raster scanning the seeds in a degassed water bath. Seed images were visually inspected and analyzed using the integrated-optical-density (IOD) method.
RESULTS: Corrugated seeds appear as contiguous objects over the range of frequencies and orientations examined, whereas standard seeds appear as contiguous objects from 90 degrees to 80 degrees only. The ranges and means of the backscattered IOD ratio of the seeds from 85 degrees to 20 degrees were: (corrugated vs. standard) 1.48 to 3.72 (2.32 +/- 0.62) for 5 MHz and 1.26 to 3.77 (2.19 +/- 0.84) for 7.5 MHz and (corrugated vs. RS) 1.21 to 9.53 (2.98 +/- 2.48) for 5 MHz and 1.008 to 10.86 (2.79 +/- 3.08) for 7.5 MHz. Backscattered signal increase ranged from 1.66 dB to 20.7 dB for the corrugated seed as compared to the other seeds.
CONCLUSIONS: Corrugated seeds produce greater backscatter signal and a more readily identifiable seed image over a large range of seed orientation as compared with standard brachytherapy seeds.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14575850     DOI: 10.1016/s0360-3016(03)01566-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  In vitro comparative study of vibro-acoustography versus pulse-echo ultrasound in imaging permanent prostate brachytherapy seeds.

Authors:  F G Mitri; B J Davis; J F Greenleaf; M Fatemi
Journal:  Ultrasonics       Date:  2008-04-29       Impact factor: 2.890

2.  Feasibility of vibro-acoustography with a quasi-2D ultrasound array transducer for detection and localizing of permanent prostate brachytherapy seeds: a pilot ex vivo study.

Authors:  Mohammad Mehrmohammadi; Azra Alizad; Randall R Kinnick; Brian J Davis; Mostafa Fatemi
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

3.  Vibro-acoustography imaging of permanent prostate brachytherapy seeds in an excised human prostate--preliminary results and technical feasibility.

Authors:  F G Mitri; B J Davis; M W Urban; A Alizad; J F Greenleaf; G H Lischer; T M Wilson; M Fatemi
Journal:  Ultrasonics       Date:  2008-10-30       Impact factor: 2.890

4.  Photoacoustic imaging of prostate brachytherapy seeds.

Authors:  Jimmy L Su; Richard R Bouchard; Andrei B Karpiouk; John D Hazle; Stanislav Y Emelianov
Journal:  Biomed Opt Express       Date:  2011-07-12       Impact factor: 3.732

5.  Intraprostatic fiducial markers: a potential application for ultrasound-guided radiotherapy in prostate cancer.

Authors:  A Vavassori; Ba Jereczek-Fossa; D Zerini; L De Cicco; R Cambria; F Cattani; C Garibaldi; M Ciocca; R Orecchia
Journal:  Ecancermedicalscience       Date:  2009-05-12

6.  Treatment planning consideration for prostate implants with the new linear RadioCoil 103Pd brachytherapy source.

Authors:  Ali S Meigooni; Shahid B Awan; Venkata Rachabatthula; Rafiq A Koona
Journal:  J Appl Clin Med Phys       Date:  2005-08-12       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.