Literature DB >> 14568714

Cl- sensitive biosensor used electrolyte-solution-gate diamond FETs.

Kwang-Soup Song1, Toshikatsu Sakai, Hirofumi Kanazawa, Yuta Araki, Hitoshi Umezawa, Minoru Tachiki, Hiroshi Kawarada.   

Abstract

We have investigated the electrolyte-solution-gate field effect transisitors (SGFETs) used hydrogen terminated (H-terminated) or partially oxygen terminated (O-terminated) polycrystalline diamond surface in the Cl- and Br- ionic solutions. The H-terminated channel SGFETs are insensitive to pH values in electrolyte solutions. The threshold voltages of the diamond SGFETs shift according to the density of Cl- and Br- ions about 30 mV/decade. One of the attractive biomedical applications for the Cl- sensitive SGFETs is the detection of chloride density in blood or in sweat especially in the case of cystic fibrosis. The sensitivities of Cl- and Br- ions have been lost on the partially O-terminated diamond surface. These phenomena can be explained by the polarity of surface change on the H-terminated and the O-terminated surface.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14568714     DOI: 10.1016/s0956-5663(03)00174-x

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  1 in total

1.  Silver-manganese nanocomposite modified screen-printed carbon electrode in the fabrication of an electrochemical, disposable biosensor strip for cystic fibrosis.

Authors:  P Arun Kumar; Aarathi Pradeep; Bipin Kumar G Nair; T G Satheesh Babu; Punathil Vasu Suneesh
Journal:  Mikrochim Acta       Date:  2022-08-11       Impact factor: 6.408