Fabric tears as a new cause of type III endoleak with Ancure endograft.

Arno Teutelink1, Maarten J van der Laan, Ross Milner, Jan D Blankensteijn.   

Abstract

PURPOSE: We present two case reports of type IIIb endoleak. One was due to fabric erosion caused by placement of a stent (Wallstent; W. L. Gore & Associates, Flagstaff, Ariz) after endovascular aneurysm repair; the other arose spontaneously. In both cases, an Ancure endograft (Guidant/EVT, Menlo Park, Calif) was placed. CASE REPORTS: In case 1, a large endoleak developed 36 months after uncomplicated endovascular treatment of an abdominal aortic aneurysm. In case 2, endoleak developed 30 months after a complicated procedure. In both cases, two Wallstents were used to treat type I endoleak and limb kinking in the first postoperative months. One type III endoleak was within the endograft at the level of the stents.
CONCLUSION: To our knowledge, these are the first type III endoleaks reported in association with Ancure endografts. Placement of Wallstents inside endografts is of concern, and another indication for close monitoring during follow-up.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14560241     DOI: 10.1016/s0741-5214(03)00416-6

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  9 in total

1.  Late type III endoleak from fabric tears of a zenith stent graft: report of a case.

Authors:  Hiroshi Banno; Hirofumi Morimae; Tsutomu Ihara; Masayoshi Kobayashi; Kiyohito Yamamoto; Kimihiro Komori
Journal:  Surg Today       Date:  2012-09-14       Impact factor: 2.549

2.  Endovascular Management of Type I Endoleak with Fenestrated Aortic "Cuff" and Afterwards Treatment of Endoleak Type III.

Authors:  Stevo Duvnjak
Journal:  Int J Angiol       Date:  2015-01-28

Review 3.  Secondary interventions following endovascular repair of abdominal aortic aneurysm.

Authors:  Naoki Toya; Yuji Kanaoka; Takao Ohki
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-10-22

4.  Delayed Type IIIb endoleak secondary to graft fabric tear 7 years following implantation of a Medtronic Talent endovascular aortic device: A case report and review of the literature.

Authors:  Vinay Kansal; Sudhir Nagpal
Journal:  SAGE Open Med Case Rep       Date:  2016-09-20

5.  Delayed Type III Endoleak Caused by Fabric Erosion after Endovascular Repair of an Abdominal Aortic Aneurysm.

Authors:  Jae Hang Lee; Eung-Joong Kim; Jin-Ho Choi
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2017-08-05

6.  Endovascular Repair for a Ruptured AAA due to a Combined Type IIIb and Ia Endoleak.

Authors:  Konstantinos Ioannis Avgerinos; Nikolaos Melas; Athanasios Saratzis; Marianthi V Tympanidou; Nikolaos Saratzis; Ioannis Lazaridis
Journal:  Case Rep Vasc Med       Date:  2018-04-24

7.  Type III B endoleak leading to aortic rupture after endovascular repair: analysis of errors in follow up and treatment.

Authors:  Marco Leopardi; Alessia Salerno; Pietro Scarpelli; Marco Ventura
Journal:  CVIR Endovasc       Date:  2018-11-06

8.  Late Type 3b Endoleak with an Endurant Endograft.

Authors:  Mehmet Barburoglu; Bulent Acunas; Yilmaz Onal; Murat Ugurlucan; Omer Ali Sayin; Ufuk Alpagut
Journal:  Case Rep Radiol       Date:  2015-12-20

9.  Aneurysm sac enlargement 16 years after endovascular aortic aneurysm repair due to late type IIIb endoleak: A case report.

Authors:  Yuri Murakami; Naoki Toya; Soichiro Fukushima; Eisaku Ito; Tadashi Akiba; Takao Ohki