Validity and accuracy of regional bioelectrical impedance devices to determine whole-body fatness.

Henry C Lukaski1, William A Siders.   

Abstract

OBJECTIVE: Growing emphasis on obesity as a risk factor for chronic diseases and commercial availability of impedance devices for the at-home assessment of body fatness have stimulated the need for a critical evaluation of the validity of these instruments. This study determined the reproducibility and accuracy of two commercial impedance devices that use upper (hand-to-hand) or lower (foot-to-foot) body contact electrode placements in adults with a wide range of body fatness.
METHODS: Body composition was assessed with dual x-ray absorptiometry in apparently healthy adults (62 women and 48 men) ages 21 to 60 y, with a range in body mass index of 18.6 to 40.5 kg/m2. Variability in body fatness predicted with the regional body impedance devices was determined in 10 adults on 5 consecutive d. A 50-kHz, tetrapolar bioelectrical impedance plethysmograph with surface electrode placements on the upper and lower limbs was used to determine reference regional and whole-body impedance values.
RESULTS: Variability in body mass (1%) over 5 d was less than body fatness predicted with the upper (2-10%) and lower (3-5%) body devices. Regional and whole-body impedance values were different (P < 0.05) in the women, whereas upper and lower body values were lower (P < 0.05) than whole-body impedance in the men. Dual x-ray absorptiometric determinations of body fatness were similar to predictions based on models derived from physical characteristics (age, stature, body mass, and sex) but significantly different (P < 0.05) from estimates from the impedance devices, which underestimated body fatness. Bias in predictions of body fatness with the regional devices was systematically (P < 0.0001) related to body fatness.
CONCLUSION: Use of regional impedance devices to assess body fatness is limited by a lack of precision and accuracy.

Entities:  

Mesh:

Year:  2003        PMID: 14559320     DOI: 10.1016/s0899-9007(03)00166-7

Source DB:  PubMed          Journal:  Nutrition        ISSN: 0899-9007            Impact factor:   4.008


  14 in total

1.  Comparison of body fat-free masses calculated from hand-to-foot and foot-to-foot resistances with DXA measurements.

Authors:  Sana Bousbiat; Michel Y Jaffrin; Edwige Dongmo
Journal:  Med Biol Eng Comput       Date:  2011-09-14       Impact factor: 2.602

2.  Narrowband diffuse reflectance spectroscopy in the 900-1000 nm wavelength region to quantify water and lipid content of turbid media.

Authors:  Jesse H Lam; Kelsey J Tu; Sehwan Kim
Journal:  Biomed Opt Express       Date:  2021-05-04       Impact factor: 3.732

3.  Development and validation of bioelectrical impedance prediction equations estimating regional lean soft tissue mass in middle-aged adults.

Authors:  Luís B Sardinha; Gil B Rosa; Megan Hetherington-Rauth; Inês R Correia; João P Magalhães; Analiza M Silva; Henry Lukaski
Journal:  Eur J Clin Nutr       Date:  2022-10-17       Impact factor: 4.884

4.  Validity of bioelectrical impedance analysis to estimate body composition changes after bariatric surgery in premenopausal morbidly women.

Authors:  Silvia Savastano; Annamaria Belfiore; Carolina Di Somma; Concetta Mauriello; Annalisa Rossi; Genoveffa Pizza; Annalba De Rosa; Giovanni Prestieri; Luigi Angrisani; Annamaria Colao
Journal:  Obes Surg       Date:  2009-10-28       Impact factor: 4.129

5.  Predicting body composition using foot-to-foot bioelectrical impedance analysis in healthy Asian individuals.

Authors:  Chun-Shien Wu; Yu-Yawn Chen; Chih-Lin Chuang; Li-Ming Chiang; Gregory B Dwyer; Ying-Lin Hsu; Ai-Chun Huang; Chung-Liang Lai; Kuen-Chang Hsieh
Journal:  Nutr J       Date:  2015-05-19       Impact factor: 3.271

6.  Hand-to-Hand Model for Bioelectrical Impedance Analysis to Estimate Fat Free Mass in a Healthy Population.

Authors:  Hsueh-Kuan Lu; Li-Ming Chiang; Yu-Yawn Chen; Chih-Lin Chuang; Kuen-Tsann Chen; Gregory B Dwyer; Ying-Lin Hsu; Chun-Hao Chen; Kuen-Chang Hsieh
Journal:  Nutrients       Date:  2016-10-21       Impact factor: 5.717

7.  A Smartphone Application for Personal Assessments of Body Composition and Phenotyping.

Authors:  Gian Luca Farina; Fabrizio Spataro; Antonino De Lorenzo; Henry Lukaski
Journal:  Sensors (Basel)       Date:  2016-12-17       Impact factor: 3.576

8.  The novel application of artificial neural network on bioelectrical impedance analysis to assess the body composition in elderly.

Authors:  Kuen-Chang Hsieh; Yu-Jen Chen; Hsueh-Kuan Lu; Ling-Chun Lee; Yong-Cheng Huang; Yu-Yawn Chen
Journal:  Nutr J       Date:  2013-02-06       Impact factor: 3.271

9.  Comparisons of a Multi-Frequency Bioelectrical Impedance Analysis to the Dual-Energy X-Ray Absorptiometry Scan in Healthy Young Adults Depending on their Physical Activity Level.

Authors:  Julien Verney; Chloé Schwartz; Saliha Amiche; Bruno Pereira; David Thivel
Journal:  J Hum Kinet       Date:  2015-10-14       Impact factor: 2.193

10.  Comparison of Standing Posture Bioelectrical Impedance Analysis with DXA for Body Composition in a Large, Healthy Chinese Population.

Authors:  Kuen-Tsann Chen; Yu-Yawn Chen; Chia-Wei Wang; Chih-Lin Chuang; Li-Ming Chiang; Chung-Liang Lai; Hsueh-Kuan Lu; Gregory B Dwyer; Shu-Ping Chao; Ming-Kuei Shih; Kuen-Chang Hsieh
Journal:  PLoS One       Date:  2016-07-28       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.