| Literature DB >> 14552826 |
Cornelia Buerger1, Christian Joukhadar, Markus Muller, Charlotte Kloft.
Abstract
Linezolid is a new, promising antibacterial agent to treat severe infections. A rapid HPLC assay using UV detection for the determination in microdialysate and human plasma was developed. After sample preparation, using acetonitrile for plasma and water for microdialysate, 20 microl was injected and separated on a RP-18 column. Overall, the assay exhibited good precision and accuracy. The diffusion properties of linezolid investigated in in vitro microdialysis experiments revealed a mean relative recovery of 77.5% (CV: 5.4%; delivery and recovery experiments). Following characterization of linezolid in in vitro microdialysis, the setting is suitable for application in clinical studies.Entities:
Mesh:
Substances:
Year: 2003 PMID: 14552826 DOI: 10.1016/j.jchromb.2003.08.019
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205