Use of a self-expanding occluding stent for nonsurgical closure of patent ductus arteriosus in dogs.

David Sisson1.   

Abstract

OBJECTIVE: To evaluate the clinical application of a catheter-delivered, self-expanding occluding stent for closure of patent ductus arteriosus (PDA) in dogs.
DESIGN: Prospective study. ANIMALS: 23 client-owned dogs weighing at least 3 kg (6.6 lb). PROCEDURE: Dogs were evaluated by physical examination, electrocardiography, thoracic radiography, and 2-dimensional, M-mode, spectral and color-flow Doppler echocardiography to confirm the diagnosis and obtain baseline measures. Shunt severity and ductal size and anatomy were established by means of angiography. With fluoroscopic guidance, the occluding stent, attached to a delivery cable, was maneuvered though the right side of the heart into the ductus via a prepositioned introducer sheath. After angiographic verification of appropriate stent placement, the delivery cable was detached, and the introducer sheath was withdrawn. Closure of the PDA was evaluated by means of angiography 15 minutes after stent deployment and by echocardiography 1 and 3 months after the procedure.
RESULTS: There were no operative deaths. There were 2 deployment failures, both attributable to avoidable operator errors. Angiography performed after stent deployment indicated PDA closure in 13 of 20 (65%) dogs. There were 2 postoperative deaths in dogs with heart failure; both deaths were thought to be unrelated to use of the occluding stent. Complete PDA closure, determined by Doppler color-flow echocardiography, was evident in 17 of 19 dogs within 3 months and in 1 additional dog within 1 year of stent deployment, resulting in closure in 18 of 19 dogs completing the study protocol. CONCLUSIONS AND CLINICAL RELEVANCE: Results suggest that a catheter-delivered occluding stent can be used successfully to close PDAs in dogs.

Entities:  

Mesh:

Year:  2003        PMID: 14552489     DOI: 10.2460/javma.2003.223.999

Source DB:  PubMed          Journal:  J Am Vet Med Assoc        ISSN: 0003-1488            Impact factor:   1.936


  4 in total

1.  An experimental canine patent ductus arteriosus occlusion device based on shape memory polymer foam in a nitinol cage.

Authors:  Mark A Wierzbicki; Sarah B Raines; Sonya G Gordon; John C Criscione; Ashley B Saunders; Scott Birch; Bradley Due; Brandis Keller; Landon D Nash; Matthew W Miller; Duncan J Maitland
Journal:  J Mech Behav Biomed Mater       Date:  2017-07-25

2.  Mechanical and in vitro evaluation of an experimental canine patent ductus arteriosus occlusion device.

Authors:  Mark A Wierzbicki; Jesse Bryant; Matthew W Miller; Brandis Keller; Duncan J Maitland
Journal:  J Mech Behav Biomed Mater       Date:  2015-12-21

3.  Late recanalization after complete occlusion of patent ductus arteriosus in a Pembroke Welsh Corgi with von Willebrand disease.

Authors:  Masaki Kochi; Keisuke Sugimoto; Michito Inoue; Takuma Aoki
Journal:  Vet Med Sci       Date:  2021-09-19

4.  Transvenous embolization of moderate to large patent ductus arteriosus in dogs using the Amplatzer vascular plug II.

Authors:  Nicolai Hildebrandt; Andreas Stosic; Estelle Henrich; Nicola Wiedemann; Gabriel Wurtinger; Matthias Schneider
Journal:  J Vet Intern Med       Date:  2021-12-16       Impact factor: 3.333

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.