Accuracy and feasibility of daily infrared tympanic membrane temperature measurements in the identification of fever in children.

Birgit K van Staaij1, Maroeska M Rovers, Anne G Schilder, Arno W Hoes.   

Abstract

BACKGROUND: Fever is an important physical sign in infectious diseases in childhood. Daily assessment of fever may be used to monitor the occurrence of infections for research purposes. The infrared tympanic membrane thermometer (ITT) seems ideal for measuring central body temperature. Its accuracy, however, is still debated, and the feasibility of daily temperature measurements with an ITT has not been described.
OBJECTIVE: To determine the accuracy and feasibility of daily infrared tympanic membrane temperature measurements at home.
METHODS: Tympanic membrane temperature was measured by the Braun Pro 3000, rectal temperature by the Omron MC 63. To establish the accuracy of the ITT, rectal and tympanic membrane temperature measurements were performed in 41 children, aged 2-10 years. To establish the feasibility of daily infrared tympanic temperature measurements at home, parents of 21 children performed daily measurements for 2 weeks.
RESULTS: With fever defined as a body temperature of > or =38.0 degrees C, sensitivity was 93.3%, specificity 92%, positive predictive value 87.5%, negative predictive value 95.8%, and the concordance rate 92.5%. Most of the parents found the instructions for the use of the ITT clear (93%) and the ITT easy to use (86%). During the test period of 2 weeks the technique of the tympanic membrane temperature measurements remained adequate in 93% of the parents.
CONCLUSION: The tympanic membrane temperature measured by the Braun Pro 3000 accurately reflects rectal temperature, validly assesses the presence of fever in children, and is easy to use. The Braun ITT therefore is an adequate tool to assess fever and may be used both in a clinical setting and for research purposes.

Entities:  

Mesh:

Year:  2003        PMID: 14550963     DOI: 10.1016/s0165-5876(03)00198-8

Source DB:  PubMed          Journal:  Int J Pediatr Otorhinolaryngol        ISSN: 0165-5876            Impact factor:   1.675


  13 in total

1.  A comparison of different methods of temperature measurement by mothers and physicians in healthy newborns.

Authors:  Mehmet Yekta Oncel; Kadir Serafettin Tekgunduz; Ramazan Ozdemir; Erhan Calisici; Sevilay Karahan; Omer Erdeve; Serife Suna Oguz; Ugur Dilmen
Journal:  Indian J Pediatr       Date:  2012-06-02       Impact factor: 1.967

Review 2.  Adenotonsillectomy for upper respiratory infections: evidence based?

Authors:  B K van Staaij; E H van den Akker; G J M G van der Heijden; A G Schilder; A W Hoes
Journal:  Arch Dis Child       Date:  2005-01       Impact factor: 3.791

3.  Effect of wearing an N95 respirator on infrared tympanic membrane temperature measurements.

Authors:  Jung-Hyun Kim; Raymond J Roberge; Jeffrey B Powell
Journal:  J Clin Monit Comput       Date:  2014-12-20       Impact factor: 2.502

4.  Effectiveness of adenotonsillectomy in children with mild symptoms of throat infections or adenotonsillar hypertrophy: open, randomised controlled trial.

Authors:  Birgit K van Staaij; Emma H van den Akker; Maroeska M Rovers; Gerrit Jan Hordijk; Arno W Hoes; Anne G M Schilder
Journal:  BMJ       Date:  2004-09-10

5.  Effectiveness of adenoidectomy in children with recurrent upper respiratory tract infections: open randomised controlled trial.

Authors:  M T A van den Aardweg; C W B Boonacker; M M Rovers; A W Hoes; A G M Schilder
Journal:  BMJ       Date:  2011-09-06

6.  Accuracy of parents in measuring body temperature with a tympanic thermometer.

Authors:  Joan L Robinson; Hsing Jou; Donald W Spady
Journal:  BMC Fam Pract       Date:  2005-01-11       Impact factor: 2.497

7.  Acceptance of temporal artery thermometry by Nigerian mothers: a comparison with the traditional methods.

Authors:  Odinaka Kelechi; Edelu Benedict; Emeka Nwolisa; Amamilo Ifeyinwa; Okolo Seline
Journal:  Pan Afr Med J       Date:  2014-11-11

8.  Evaluation of a Wearable Non-Invasive Thermometer for Monitoring Ear Canal Temperature during Physically Demanding (Outdoor) Work.

Authors:  Charlotte Christina Roossien; Audy Paul Hodselmans; Ronald Heus; Michiel Felix Reneman; Gijsbertus Jacob Verkerke
Journal:  Int J Environ Res Public Health       Date:  2021-05-04       Impact factor: 3.390

9.  The Effect of Various Hot Environments on Physiological Responses and Information Processing Performance Following Firefighting Activities in a Smoke-Diving Room.

Authors:  Rasoul Hemmatjo; Majid Motamedzade; Mohsen Aliabadi; Omid Kalatpour; Maryam Farhadian
Journal:  Saf Health Work       Date:  2017-02-27

10.  Mass screening of suspected febrile patients with remote-sensing infrared thermography: alarm temperature and optimal distance.

Authors:  Ming-Fu Chiang; Po-Wei Lin; Li-Fong Lin; Hung-Yi Chiou; Ching-Wen Chien; Shu-Fen Chu; Wen-Ta Chiu
Journal:  J Formos Med Assoc       Date:  2008-12       Impact factor: 3.282

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.