Preliminary evaluation of a fuzzy logic-based automatic quantitative analysis in myocardial SPECT.

Florent Cachin1, Janusz Lipiecki, Danièle Mestas, Aimé Amonchot, Benjamin Geissler, Cyril Thouly, Jean Ponsonnaille, Jean Maublant.   

Abstract

UNLABELLED: This study validates a new quantitative myocardial perfusion SPECT software.
METHODS: The processing starts with the extraction of the morphologic skeleton of the left ventricular myocardium from reconstructed transverse sections. Fuzzy logic is used to decide whether a pixel belongs to the myocardium and any perfusion defect is filled according to a truncated bullet model. The resulting image is partitioned in 18 isovolumetric sectors. Sex-matched normal limits, criteria of abnormality for rest (201)Tl and (99m)Tc-labeled perfusion tracers, reproducibility studies, and detection of coronary artery disease were developed and validated in an overall population of 343 patients. The sex- and tracer-matched means and SDs of a normal response were calculated in 93 male and 93 female patients with a <5% likelihood of coronary artery disease. Reproducibility measurements and assignment of different sectors of the myocardium to a specific coronary were performed from data collected in 49 and 60 patients, respectively. The accuracy of the detection of a coronary artery occlusion was assessed in 48 patients who also underwent coronary angiography.
RESULTS: The intra- and interoperator reproducibility of the sectorial activity was high with a linear regression coefficient of 0.97 and a SD of the difference measurement at 4.4% and 3.8%, respectively. Overall sensitivity and specificity for the detection of occluded coronary artery were 90% and 80%, respectively. For the detection of left anterior descending, left circumflex, and right artery coronary occlusion, sensitivity was 92%, 75%, and 92.5%, respectively, and specificity was 75%, 78%, and 90%, respectively.
CONCLUSION: The new quantitative myocardial perfusion SPECT software appears to be a very helpful program for the objective analysis of perfusion tracer distribution in myocardial SPECT and a very accurate tool in the detection and localization of coronary artery occlusion.

Entities:  

Mesh:

Year:  2003        PMID: 14530477

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  2 in total

1.  Influence of infarct-zone viability detected by rest Tc-99m sestamibi gated SPECT on left ventricular remodeling after acute myocardial infarction treated by percutaneous transluminal coronary angioplasty in the acute phase.

Authors:  Janusz Lipiecki; Florent Cachin; Nicolas Durel; Olivier de Tauriac; Jean Ponsonnaille; Jean Maublant
Journal:  J Nucl Cardiol       Date:  2004 Nov-Dec       Impact factor: 5.952

2.  Fuzzy Rule-Based Classification System for Assessing Coronary Artery Disease.

Authors:  Reza Ali Mohammadpour; Seyed Mohammad Abedi; Somayeh Bagheri; Ali Ghaemian
Journal:  Comput Math Methods Med       Date:  2015-09-13       Impact factor: 2.238

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.