Pia Haraldsson1, Tommy Knöös, Håkan Nyström, Per Engström.
Abstract
Thermoluminescent dosimeters (TLDs) are used for verification of the delivered dose during IMRT treatment of head and neck carcinomas. The TLDs are put into a plastic tube, which is placed in the nasal cavities through the treated volume. In this study, the dose distribution to a phantom having a cylindrical air cavity containing a tube was calculated by Monte Carlo methods and the results were compared with data from a treatment planning system (TPS) to evaluate the accuracy of the TLD measurements. The phantom was defined in the DOSXYZnrc Monte Carlo code and calculations were performed with 6 MV fields, with the TLD tube placed at different positions within the cylindrical air cavity. A similar phantom was defined in the pencil beam based TPS. Differences between the Monte Carlo and the TPS calculations of the absorbed dose to the TLD tube were found to be small for an open symmetrical field. For a half-beam field through the air cavity, there was a larger discrepancy. Furthermore, dose profiles through the cylindrical air cavity show, as expected, that the treatment planning system overestimates the absorbed dose in the air cavity. This study shows that when using an open symmetrical field, Monte Carlo calculations of absorbed doses to a TLD tube in a cylindrical air cavity give results comparable to a pencil beam based treatment planning system.Entities: