Albuterol aerosol delivered via metered-dose inhaler to intubated pediatric models of 3 ages, with 4 spacer designs.

Piush Mandhane1, Peter Zuberbuhler, Carlos F Lange, Warren H Finlay.   

Abstract

OBJECTIVE: To determine the amount of albuterol, in various particle size ranges, delivered from a hydrofluoroalkane-propelled metered-dose inhaler (Airomir) in 3 models of pediatric intubation (ages 8 months, 4 years, and 16 years) using 4 types of aerosol reservoir: 3 spacers (ACE, AeroChamber HC MV, metal NebuChamber without 1-way valve) and 1 holding chamber (metal NebuChamber with 1-way valve).
METHODS: Five reservoirs of each type were tested with albuterol sulfate delivered via metered-dose inhaler that delivers 100 microg of albuterol per actuation. Each reservoir was connected to an endotracheal tube (ETT) that corresponded to the given patient age (8 months = 4 French; 4 years = 5 French; 16 years = 7.5 French) and to a valved system that allowed connection of the ETT to a cascade impactor. Simulated tidal volumes representative of children of the given ages were passed through the reservoir. Both the cascade impactor and the ETT were enclosed within a 100% humidity, 37 degrees C environment.
RESULTS: For the total amount of albuterol inhaled onto the impactor, and both the 1.1-4.7 microm and 1.1-3.3 microm inhaled fine-particle fractions, the NebuChamber-with-valve showed significantly greater drug delivery than the NebuChamber-without-valve, the AeroChamber HC MV, or the ACE (p < 0.001). Among the reservoirs without valves the NebuChamber showed significantly greater delivery than the AeroChamber HC MV or ACE (p < 0.001) for total drug deposition and for both the 1.1-4.7 microm and 1.1-3.3 microm fine-particle fractions. These results were consistent over all age groups. The AeroChamber HC MV had significantly greater delivery (total deposition) than the ACE (p < 0.001), except in the 4-year-old model. There were no significant differences between the AeroChamber HC MV and the ACE for either the 1.1-4.7 microm or the 1.1-3.3 microm fine-particle fraction.
CONCLUSION: An aerosol reservoir with 1-way valve positioned between the spacer and the ETT improved the amount of inhaled albuterol 300-900%, compared to the other reservoirs.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14525631

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  1 in total

Review 1.  In vitro and in vivo aspects of cascade impactor tests and inhaler performance: a review.

Authors:  Jolyon Mitchell; Steve Newman; Hak-Kim Chan
Journal:  AAPS PharmSciTech       Date:  2007-12-21       Impact factor: 3.246

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.