| Literature DB >> 14522043 |
Dafeng Chu1, Jingkai Gu, Wanhui Liu, J Paul Fawcett, Qingguang Dong.
Abstract
A high-performance liquid chromatographic (HPLC) method has been developed for the simultaneous determination of 5-fluorouracil (5-FU) and its prodrug, tegafur (TF), in dog plasma. 5-FU, the internal standard, 5-bromouracil, and TF were separated on a C18 Spherisorb ODS2 column using isocratic elution with retention times of 4.4, 8.0 and 21.2 min, respectively. Detection by UV absorption at 260 nm gave a limit of quantitation of 4 microg/l for 5-FU in plasma. Calibration curves for 5-FU and TF were linear over the ranges of 4-160 microg/l and 0.48-19.2 mg/l, respectively. Intra- and inter-day precision over these concentration ranges were <10.9 and <13.6% for 5-FU and TF, respectively, with good accuracy for both compounds. The method was successfully applied to define plasma concentration-time curves of TF and 5-FU in dogs administered a single oral dose containing TF (100 mg) and uracil (224 mg).Entities:
Mesh:
Substances:
Year: 2003 PMID: 14522043 DOI: 10.1016/s1570-0232(03)00571-3
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205